TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
Published September 24, 2014 · Recall initiated: August 28, 2014
Summary
FDA records medical device Class II recall Z-2657-2014, published September 24, 2014 and initiated August 28, 2014. Recalling firm: Motion Concepts; Status: Terminated; Reported quantity: US: 59 units; Lot: All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Motion Concepts
- Distributor
- US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.
- Lot
- All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Canada
Related recalls
- FDAZenith Slide W-I-D-E Long Term Care Bed, Model(s) APS98674, APS9867406, APS986741, APS986742, ZG58674,September 24, 2014
- FDAZOLL Pro-padz Sterile Adult Multi-Function Electrodes Part number: 8900-4012 These MFE Electrodes are used inSeptember 24, 2014
- FDAB.Braun Accuflo 100ML/H 100ML Elastomeric Infusion Device. AccuFlo devices are intended for patients requiringSeptember 24, 2014
- FDAZOLL Pro-padz Sterile Adult Multi-Function Electrodes ((singles) Part number: 8900-4013 These MFE ElectrodesSeptember 24, 2014
- FDA4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatmentSeptember 24, 2014
- FDAZOLL Pro-padz Sterile Adult Multi-Function Electrodes with 10-ft (3 m) leadwires Part number: 8900-4052 TheseSeptember 24, 2014
Questions about this recall
Was TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System. recalled?
Yes. FDA published recall Z-2657-2014 on September 24, 2014. Reported hazard: Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
What should I do if I have this product?
Voluntary: Firm initiated · Letter