MAGNETOM Verio. Model Number: 10276755.
Published October 8, 2025 · Recall initiated: August 28, 2025
Summary
FDA records medical device Class I recall Z-2664-2025, published October 8, 2025 and initiated August 28, 2025. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Ongoing; Reported quantity: 154 units; Lot: Model Number: 10276755. UDI Numbers: N/A. Serial Numbers: 40649, 40585, 40781, 40243, 40408, 40628, 40848, 40870, 40783, 40473, 40240, 40497, 40564, 40425, 40778, 40389, 40340, 40523, 40582, 40168, 40291, 40283, 40201, 40850, 40331, 40126, 40744, 40821, 40433, 40502, 40179, 40246, 40273, 40766, 40753, 40828, 40737, 40352, 40313, 40382, 40670, 40387, 40631, 40669, 40131, 40541, 40466, 40292, 40375, 40161, 40745, 40268, 40284, 40325, 40799, 40190, 40877, 40104, 40195, 40407, 40640, 40797, 40254, 40142, 40665, 40622, 40264, 40420, 40633, 40150, 40514, 40727, 40710, 40328, 40156, 40800, 40692, 40165, 40441, 40864, 40482, 40472, 40859, 40220, 40858, 40779, 40346, 40295, 40233, 40629, 40681, 40464, 40568, 40200, 40663, 40334, 40277, 40276, 40879, 40222, 40361, 40553, 40378, 40714, 40715, 40252, 40250, 40423, 40204, 40486, 40251, 40488, 40109, 40855, 40429, 40232, 40157, 40440, 40259, 40545, 40556, 40687, 40140, 40237, 40399, 40312, 40235, 40697, 40684, 40398, 40871, 40700, 40368, 40656, 40606, 40163, 40278, 40562, 40671, 40526, 40658, 40644, 40376, 40686, 40713, 40775, 40863, 40326, 40409, 40116. Additional serial number added 11/14/2025: 40250. Additional serial number added 01/21/2026: 40459. Additional serial number added 2/25/2026: 40464. Additional serial number added 4/17/2026: 40711. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
- Lot
- Model Number: 10276755. UDI Numbers: N/A. Serial Numbers: 40649, 40585, 40781, 40243, 40408, 40628, 40848, 40870, 40783, 40473, 40240, 40497, 40564, 40425, 40778, 40389, 40340, 40523, 40582, 40168, 40291, 40283, 40201, 40850, 40331, 40126, 40744, 40821, 40433, 40502, 40179, 40246, 40273, 40766, 40753, 40828, 40737, 40352, 40313, 40382, 40670, 40387, 40631, 40669, 40131, 40541, 40466, 40292, 40375, 40161, 40745, 40268, 40284, 40325, 40799, 40190, 40877, 40104, 40195, 40407, 40640, 40797, 40254, 40142, 40665, 40622, 40264, 40420, 40633, 40150, 40514, 40727, 40710, 40328, 40156, 40800, 40692, 40165, 40441, 40864, 40482, 40472, 40859, 40220, 40858, 40779, 40346, 40295, 40233, 40629, 40681, 40464, 40568, 40200, 40663, 40334, 40277, 40276, 40879, 40222, 40361, 40553, 40378, 40714, 40715, 40252, 40250, 40423, 40204, 40486, 40251, 40488, 40109, 40855, 40429, 40232, 40157, 40440, 40259, 40545, 40556, 40687, 40140, 40237, 40399, 40312, 40235, 40697, 40684, 40398, 40871, 40700, 40368, 40656, 40606, 40163, 40278, 40562, 40671, 40526, 40658, 40644, 40376, 40686, 40713, 40775, 40863, 40326, 40409, 40116. Additional serial number added 11/14/2025: 40250. Additional serial number added 01/21/2026: 40459. Additional serial number added 2/25/2026: 40464. Additional serial number added 4/17/2026: 40711;
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAMAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253,October 8, 2025
- FDAArgus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006October 8, 2025
Questions about this recall
Was MAGNETOM Verio. Model Number: 10276755. recalled?
Yes. FDA published recall Z-2664-2025 on October 8, 2025. Reported hazard: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
What should I do if I have this product?
Voluntary: Firm initiated · Letter