Back to search
FDAFDA · Food and drugs·Recall number Z-2670-2017

Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.

Published July 12, 2017 · Recall initiated: April 4, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro¿ needles, Smiths Medical part numbers EL 1915 and SB5021.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
US Distribution to the state of : IN
Lot
3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.