FDAFDA · Food and drugs·Recall number Z-2670-2017

Jelco, Hypodermic needle Pro® needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.

Published July 12, 2017 · Recall initiated: April 4, 2017

Summary

FDA records medical device Class II recall Z-2670-2017, published July 12, 2017 and initiated April 4, 2017. Recalling firm: Smiths Medical ASD Inc.; Status: Terminated; Reported quantity: 736,000; Lot: 3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro® needles, Smiths Medical part numbers EL 1915 and SB5021.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
US Distribution to the state of : IN
Lot
3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Jelco, Hypodermic needle Pro® needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO. recalled?

Yes. FDA published recall Z-2670-2017 on July 12, 2017. Reported hazard: Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro® needles, Smiths Medical part numbers EL 1915 and SB5021.

What should I do if I have this product?

Voluntary: Firm initiated · Letter