Jelco, Hypodermic needle Pro® needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.
Published July 12, 2017 · Recall initiated: April 4, 2017
Summary
FDA records medical device Class II recall Z-2670-2017, published July 12, 2017 and initiated April 4, 2017. Recalling firm: Smiths Medical ASD Inc.; Status: Terminated; Reported quantity: 736,000; Lot: 3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro® needles, Smiths Medical part numbers EL 1915 and SB5021.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smiths Medical ASD Inc.
- Distributor
- US Distribution to the state of : IN
- Lot
- 3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of theNovember 6, 2024
- FDAENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28October 30, 2024
- FDACADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JPOctober 30, 2024
- FDACADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24October 30, 2024
- FDAElectrophysiology catheter cables are designed as electrode cables with a multi-pin connector on the distalJuly 12, 2017
- FDABoston Scientific Extractor Pro RX-S Retrieval Balloon Catheter 12-15mm Below UPN: M00547340July 12, 2017
Questions about this recall
Was Jelco, Hypodermic needle Pro® needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO. recalled?
Yes. FDA published recall Z-2670-2017 on July 12, 2017. Reported hazard: Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro® needles, Smiths Medical part numbers EL 1915 and SB5021.
What should I do if I have this product?
Voluntary: Firm initiated · Letter