ABIOMED Impella RP Flex with SmartAssist Set, Pump Set Code 1000323, containing Kit, 23Fr Introducer, 11cm, Sterile, 23 Fr, 11 cm, Product Code 2000342. heart pump and accessories
Published July 22, 2026 · Recall initiated: June 11, 2026
Summary
FDA records medical device Class I recall Z-2681-2026, published July 22, 2026 and initiated June 11, 2026. Recalling firm: Abiomed, Inc.; Status: Ongoing; Reported quantity: 5,017 units; Lot: Pump Set Code 1000323; introducer Product Code 2000342. GTIN 00813502012811. All current batches in field. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abiomed, Inc.
- Distributor
- Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
- Lot
- Pump Set Code 1000323; introducer Product Code 2000342. GTIN 00813502012811. All current batches in field.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAImpella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, aNovember 4, 2026
- FDAImpella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, aNovember 4, 2026
- FDAAutomated Impella Controller, Packaged. Primary user control interface for the Impella Catheter,November 4, 2026
- FDAAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1.September 30, 2026
- FDAAbiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with InfusionJuly 22, 2026
- FDAABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm,July 22, 2026
Questions about this recall
Was ABIOMED Impella RP Flex with SmartAssist Set, Pump Set Code 1000323, containing Kit, 23Fr Introducer, 11cm, Sterile, 23 Fr, 11 cm, Product Code 2000342. heart pump and accessories recalled?
Yes. FDA published recall Z-2681-2026 on July 22, 2026. Reported hazard: A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
What should I do if I have this product?
Voluntary: Firm initiated · Letter