DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Published August 28, 2024 · Recall initiated: July 10, 2024
Summary
FDA records medical device Class II recall Z-2691-2024, published August 28, 2024 and initiated July 10, 2024. Recalling firm: Beckman Coulter Mishima K.K.; Status: Ongoing; Reported quantity: 24 analyzers; Lot: Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Mishima K.K.
- Distributor
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.
- Lot
- Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Related recalls
- FDASystems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code:October 7, 2026
- FDACHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281March 18, 2026
- FDADisposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5AAugust 28, 2024
- FDAIBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment ofAugust 28, 2024
- FDABarricaid Annular Closure Impactor, REF: BAR-IMPACTOR, P/N: 400906-AAugust 28, 2024
- FDADxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytesAugust 7, 2024
Questions about this recall
Was DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. recalled?
Yes. FDA published recall Z-2691-2024 on August 28, 2024. Reported hazard: Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter