Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.
Published October 1, 2014 · Recall initiated: July 28, 2014
Summary
FDA records medical device Class II recall Z-2693-2014, published October 1, 2014 and initiated July 28, 2014. Recalling firm: ARJOHUNTLEIGH POLSKA Sp. z.o.o.; Status: Terminated; Reported quantity: 18 Lifts; Lot: Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ARJOHUNTLEIGH POLSKA Sp. z.o.o.
- Distributor
- Worldwide Distribution - US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.
- Lot
- Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Poland
Related recalls
- FDACarevo, Height-adjustable shower trolley, Model Number BAC1111-01November 4, 2026
- FDAArjo Tenor mobile passive patient lift, Model Numbers KHA1000 US and KHA1010 USApril 22, 2026
- FDAArjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled withJanuary 3, 2024
- FDAArjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled withJanuary 3, 2024
- FDAABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent withOctober 1, 2014
- FDAeye-pak 7407 Tray Support Cover REF 8065740745October 1, 2014
Questions about this recall
Was Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair. recalled?
Yes. FDA published recall Z-2693-2014 on October 1, 2014. Reported hazard: ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp
What should I do if I have this product?
Voluntary: Firm initiated · Letter