Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Published September 4, 2024 · Recall initiated: June 2, 2024

Summary

FDA records medical device Class II recall Z-2718-2024, published September 4, 2024 and initiated June 2, 2024. Recalling firm: Treace Medical Concepts, Inc.; Status: Ongoing; Reported quantity: 40 kits containing 40 plates; Lot: UDI: (01)00810111222464(17)290424(10)300542404. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Treace Medical Concepts, Inc.
Distributor
US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.
Lot
UDI: (01)00810111222464(17)290424(10)300542404
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component. recalled?

Yes. FDA published recall Z-2718-2024 on September 4, 2024. Reported hazard: Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone