Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.
Published September 4, 2024 · Recall initiated: June 2, 2024
Summary
FDA records medical device Class II recall Z-2718-2024, published September 4, 2024 and initiated June 2, 2024. Recalling firm: Treace Medical Concepts, Inc.; Status: Ongoing; Reported quantity: 40 kits containing 40 plates; Lot: UDI: (01)00810111222464(17)290424(10)300542404. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Treace Medical Concepts, Inc.
- Distributor
- US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.
- Lot
- UDI: (01)00810111222464(17)290424(10)300542404
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose MonitoringSeptember 4, 2024
- FDACobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEOSeptember 4, 2024
- FDAMedline Convenience kits, labeled as: 1) NERVE STIMULATOR, Pack Number CDS981579O ; 2) DR DIULUS INJECTIONSeptember 4, 2024
- FDAIon Endoluminal System, REF: 380748-65September 4, 2024
- FDAMedline convenience kits, labeled as: 1) PAIN SERVICES CDS , Pack Number CDS985445B ; 2) PAIN SERVICES CDS ,September 4, 2024
- FDAStryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following manner: drilling,September 4, 2024
Questions about this recall
Was Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component. recalled?
Yes. FDA published recall Z-2718-2024 on September 4, 2024. Reported hazard: Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone