da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
Published July 22, 2026 · Recall initiated: May 12, 2025
Summary
FDA records medical device Class II recall Z-2722-2026, published July 22, 2026 and initiated May 12, 2025. Recalling firm: Intuitive Surgical, Inc.; Status: Ongoing; Reported quantity: 11113; Lot: UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, Japan, South Korea, Hong Kong, Taiwan, France, Ireland, Spain.
- Lot
- UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.July 22, 2026
- FDAICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusionJuly 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDABrand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number:May 13, 2026
Questions about this recall
Was da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15 recalled?
Yes. FDA published recall Z-2722-2026 on July 22, 2026. Reported hazard: Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
What should I do if I have this product?
Voluntary: Firm initiated