Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Published October 8, 2014 · Recall initiated: May 30, 2014
Summary
FDA records medical device Class II recall Z-2725-2014, published October 8, 2014 and initiated May 30, 2014. Recalling firm: Instrumed International, Inc.; Status: Terminated; Reported quantity: 83 instruments; Lot: 1) Instrumed Item No.: 250-08930; Pilling Weck Inc. / Teleflex Medical Catalog Code: 728163; Lot Number: 66745. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Instrumed International, Inc.
- Distributor
- Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
- Lot
- 1) Instrumed Item No.: 250-08930; Pilling Weck Inc. / Teleflex Medical Catalog Code: 728163; Lot Number: 66745
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAD'errico perforator drills are bone cutting and drilling instruments that are used without a power source on aDecember 17, 2014
- FDAA zipser clamp is an instrument used to compress reduce blood flow to the penis and reduce pain and is soldDecember 17, 2014
- FDAMcKenzie enlarging burrs are bone cutting and drilling instruments that are used without a power source on aDecember 17, 2014
- FDAGardner-Wells traction tongs are skull tongs for traction used to immobilize a patient with a cervical spineDecember 17, 2014
- FDAReusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices mustOctober 8, 2014
- FDAReusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices mustOctober 8, 2014
Questions about this recall
Was Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue. recalled?
Yes. FDA published recall Z-2725-2014 on October 8, 2014. Reported hazard: Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
What should I do if I have this product?
Voluntary: Firm initiated · Letter