BD 10mL Syringe Luer-Lok Tip Catalog number 301997 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
Published July 26, 2017 · Recall initiated: September 16, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- Nationwide Distribution
- Lot
- 6061746 2021-02 6082982 2021-02 6089821 2021-03 6117858 2021-04 6117863 2021-04 6117889 2021-04 6117891 2021-04 6146962 2021-04 6173697 2021-05 6173767 2021-06 6173779 2021-06 6208614 2021-07
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States