Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.
Published August 12, 2020 · Recall initiated: July 6, 2020
Summary
FDA records medical device Class II recall Z-2765-2020, published August 12, 2020 and initiated July 6, 2020. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Lot: Outer Package UPN #: M00562402 Inner Package UPN #: M00562400 GTIN: 08714729501640 Lot #s: 25111007, 25133479, 25142120, 25299896, 25351474, 25357539, 25402321 Expiration Dates: 28-Jan-23 through 24-Mar- 23. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- World wide distribution.
- Lot
- Outer Package UPN #: M00562402 Inner Package UPN #: M00562400 GTIN: 08714729501640 Lot #s: 25111007, 25133479, 25142120, 25299896, 25351474, 25357539, 25402321 Expiration Dates: 28-Jan-23 through 24-Mar- 23
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDACaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop whichAugust 12, 2020
- FDABOSS DeBakey "Gator-Grip" Needle Holder 2 mm tip, 9", Part Numbers 50-2503 and 51-2503 - Product Usage:August 12, 2020
Questions about this recall
Was Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop. recalled?
Yes. FDA published recall Z-2765-2020 on August 12, 2020. Reported hazard: BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter