Back to search
FDAFDA · Food and drugs·Recall number Z-2769-2024

Centurion procedure kits labeled as: UMBILICAL VESSEL TRAY, Pack Number UVT1100

Published September 4, 2024 · Recall initiated: April 8, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
Distributor
worldwide
Lot
UVT1100, Lot Number 2021032280; UVT1100, Lot Number 2021072680; UVT1100, Lot Number 2021092780; UVT1100, Lot Number 2022011080
Risk classification
Class II
Area (FDA)
device
Country
United States