VITEK 2 Gram Negative test kits containing colistin (cs01n)

Published August 2, 2017 · Recall initiated: May 16, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Biomerieux Inc
Distributor
Domestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.
Lot
REF # (Description): 418114 (AST-GN94), 22258 (AST-N102), 412605 (AST-N194), 412921 (AST-N199), 412862 (AST-N201), 412863 (AST-N202), 412924 (AST-N205), 412918 (AST-N210), 413061 (AST-N212), 413083 (AST-N222), 413112 (AST-N225), 413171 (AST-N236), 413172 (AST-N237), 413204 (AST-N239), 413205 (AST-N240), 413391 (AST-N242), 413394 (AST-N245), 413397 (AST-N248), 413572 (AST-N249), 413722 (AST-N253), 413724 (AST-N255), 413725 (AST-N256), 413753 (AST-N261), 414164 (AST-N272), 414286 (AST-N276), 414492 (AST-N279), 414531 (AST-N280), 414532 (AST-N281), 414972 (AST-N288), 415063 (AST-N292), 415369 (AST-N293), 415433 (AST-N295), 416005 (AST-N299), 416241 (AST-N300), 416590 (AST-N303), 416913 (AST-N308), 416914 (AST-N309), 417426 (AST-N315), 417952 (AST-N318), 418204 (AST-N320), 418489 (AST-N322), 418513 (AST-N235), 418514 (AST-N326), 418630 (AST-N328), 418631 (AST-N329), 418675 (AST-N331), 418676 (AST-N332), 418985 (AST-N335), 419077 (AST-N338), 419341 (AST-N339), 420063 (AST-N343), 420440 (AST-N344), 420867 (AST-N347), 420856 (AST-N348), 421258 (AST-N352), 421297 (AST-N353), 421350 (AST-N354), 421351 (AST-N355, 421451 (AST-N357), 421441 (AST-N358), 421583 (AST-N360), 421584 (AST-N361), 421585 (AST-N362), 421692 (AST-363), 421693 (AST-N364), 421830 (AST-N365), 421853 (AST-N366), 421854 (AST-N367), 421855 (AST-N368), 410025 (AST-XN01), 413230 (AST-XN05)
Risk classification
Class II
Area (FDA)
device
Country
United States