FDAFDA · Food and drugs·Recall number Z-2776-2020

Verigene Processor SP with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.

Published August 19, 2020 · Recall initiated: June 15, 2020

Summary

FDA records medical device Class II recall Z-2776-2020, published August 19, 2020 and initiated June 15, 2020. Recalling firm: Luminex Corporation; Status: Terminated; Reported quantity: Total = 3,016 systems (2240 Verigene Processor SP and 776 Verigene Processor SP Refurbished); Lot: Verigene Processor SP, 10-0000-07, GTIN 00840487101483, Serial Numbers: 902, 907, 908, 909, 911, 912, 9098001, 9098002, 9098003, 9098005, 9098006, 9100009, 9100011, 9100012, 9100014, 9121015, 9121017, 9121018, 9121019, 9121021, 9121022, 9135024, 9135026, 9135027, 9135028, 9135029, 9163034, 9163035, 9163038, 9163039, 9163040, 9163041, 9163042, 9163043, 9163044, 9163045, 9163046, 9202047, 9202050, 9202051, 9202052, 9202053, 9202054, 9210055, 9210056, 9210057, 9210058, 9210059, 9210061, 9210062, 9230063, 9230064, 9230065, 9230066, 9230067, 9230068, 9230069, 9230070, 9230074, 9230075, 9255071, 9255072, 9255077, 9255079, 9255080, 9270080, 9270082, 9270084, 9270085, 9270086, 9270087, 9277088, 9277089, 9277090, 9277091, 9277092, 9277093, 9277094, 9288095, 9288096, 9288097, 9288102, 9293103, 9293106, 9293109, 9293111, 9298112, 9298116, 9298117, 9298118, 9313119, 9313120, 9313121, 9313122, 9313123, 9313124, 9320127, 9325138, 9325151, 9325152, 9336144, 9336145, 9336146, 9336147, 9336148, 9336149, 9336150, 9336155, 9336156, 9336157, 9336158, 9350159, 9350160, 9350161, 9350163, 9350164, 9350165, 10014001, 10014003, 10014012, 10035017, 10035105, 10035106, 10043023, 10043024, 10043025, 10043027, 10099029, 10099032, 10099033, 10099037, 10102046, 10110047, 10110048, 10110050, 10110051, 10110052, 10110053, 10110055, 10110056, 10110057, 10110058, 10110060, 10110061, 10110062, 10119063, 10119064, 10119065, 10119066, 10139080, 10139081, 10139082, 10139083, 10139084, 10162085, 10162092, 10162093, 10162094, 10182095, 10182096, 10182100, 10182101, 10182102, 10182103, 10182104, 10210105, 10210106, 10210107, 10210108, 10210109, 10210110, 10210111, 10210115, 10250117, 10250118, 10250119, 10250120, 10250121, 10250122, 10250123, 10276131, 10278128, 10278131, 10278132, 10278133, 10278163, 10295135, 10295136, 10295137, 10295138, 10295139, 10295140, 10295141, 10295142, 10295143, 10295144, 10295145, 10295146, 10295147, 10295149, 11021003, 11021008, 11101027, 11101028, 11122033, 11122034, 11152041, 11152042, 11152043, 11152044, 11152045, 11152046, 11152047, 11152048, 11152049, 11152050, 11152051, 11152052, 11152053, 11186054, 11186055, 11186056, 11186061, 11186062, 11186063, 11186064, 11186066, 11213071, 11213072, 11213073, 11213074, 11213075, 11213076, 11213078, 11213079, 11213080, 11213084, 11213085, 11236087, 11236089, 11236090, 11236091, 11236092, 11236093, 11250094, 11250095, 11250096, 11250098, 11250099, 11250100, 11250101, 11250103, 11250105, 11250106, 11250108, 11250109, 11276111, 11276112, 11276113, 11276114, 11276121, 11276122, 11276123, 11276124, 11276125, 11276127, 11276128, 11276129, 11306130, 11306131, 11306132, 11306133, 11306135, 11306136, 11306137, 11306141, 11306142, 11306144, 11306145, 11306146, 11306147, 11306148, 11306149, 11339151, 11339152, 11339153, 11339156, 11339157, 11339158, 11339159, 11339162, 11339163, 11339164, 12016001, 12016002, 12016003, 12016004, 12038011, 12038013, 12038014, 12038015, 12038016, 12038018, 12045021, 12045022, 12045023, 12045026, 12045027, 12045028, 12045029, 12045030, 12053034, 12053035, 12053036, 12053038, 12053039, 12053040, 12061041, 12061042, 12061043, 12061045, 12061048, 12061050, 12069052, 12069055, 12069056, 12069057, 12069058, 12069059, 12069060, 12079061, 12079062, 12079063, 12079064, 12079065, 12079066, 12079068, 12079069, 12079070, 12086071, 12086072, 12086073, 12086075, 12086076, 12086077, 12086078, 12086079, 12086080, 12088081, 12088082, 12088083, 12088084, 12088085, 12088086, 12088087, 12088088, 12088089, 12107091, 12107092, 12107093, 12107094, 12107095, 12107097, 12107098, 12111108, 12111109, 12111110, 12111111, 12111112, 12122056, 12128118, 12128119, 12128122, 12128123, 12128124, 12136125, 12136126, 12136127, 12136128, 12136134, 12136135, 12136136, 12142137, 12142138, 12142141, 12142142, 12145143, 12145144, 12145145, 12145146, 12145147, 12145148, 12145149, 12145150, 12145151, 12145152, 12145153, 12145154, 12156155, 12156156, 12156158, 12156160, 12156161, 12156162, 12156163, 12156164, 12156165,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Luminex Corporation
Distributor
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CN, CT, DE, DC, FL,GA, ID, IN, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NB, NV, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN TX, UT, VT, VA, WA, WV, WY, and WI. The countries of Austria, Belgium, Bulgaria, Canada, Chile, China, France, Germany, Greece, Hong Kong, Italy, Japan, Republic of Korea, Kuwait, Macau, Mexico, Netherlands, New Zealand, Republic of Russia, Saudi Arabia, Singapore, Spain, Thailand, Turkey, United Arab Emirate, and United Kingdom.
Lot
Verigene Processor SP, 10-0000-07, GTIN 00840487101483, Serial Numbers: 902, 907, 908, 909, 911, 912, 9098001, 9098002, 9098003, 9098005, 9098006, 9100009, 9100011, 9100012, 9100014, 9121015, 9121017, 9121018, 9121019, 9121021, 9121022, 9135024, 9135026, 9135027, 9135028, 9135029, 9163034, 9163035, 9163038, 9163039, 9163040, 9163041, 9163042, 9163043, 9163044, 9163045, 9163046, 9202047, 9202050, 9202051, 9202052, 9202053, 9202054, 9210055, 9210056, 9210057, 9210058, 9210059, 9210061, 9210062, 9230063, 9230064, 9230065, 9230066, 9230067, 9230068, 9230069, 9230070, 9230074, 9230075, 9255071, 9255072, 9255077, 9255079, 9255080, 9270080, 9270082, 9270084, 9270085, 9270086, 9270087, 9277088, 9277089, 9277090, 9277091, 9277092, 9277093, 9277094, 9288095, 9288096, 9288097, 9288102, 9293103, 9293106, 9293109, 9293111, 9298112, 9298116, 9298117, 9298118, 9313119, 9313120, 9313121, 9313122, 9313123, 9313124, 9320127, 9325138, 9325151, 9325152, 9336144, 9336145, 9336146, 9336147, 9336148, 9336149, 9336150, 9336155, 9336156, 9336157, 9336158, 9350159, 9350160, 9350161, 9350163, 9350164, 9350165, 10014001, 10014003, 10014012, 10035017, 10035105, 10035106, 10043023, 10043024, 10043025, 10043027, 10099029, 10099032, 10099033, 10099037, 10102046, 10110047, 10110048, 10110050, 10110051, 10110052, 10110053, 10110055, 10110056, 10110057, 10110058, 10110060, 10110061, 10110062, 10119063, 10119064, 10119065, 10119066, 10139080, 10139081, 10139082, 10139083, 10139084, 10162085, 10162092, 10162093, 10162094, 10182095, 10182096, 10182100, 10182101, 10182102, 10182103, 10182104, 10210105, 10210106, 10210107, 10210108, 10210109, 10210110, 10210111, 10210115, 10250117, 10250118, 10250119, 10250120, 10250121, 10250122, 10250123, 10276131, 10278128, 10278131, 10278132, 10278133, 10278163, 10295135, 10295136, 10295137, 10295138, 10295139, 10295140, 10295141, 10295142, 10295143, 10295144, 10295145, 10295146, 10295147, 10295149, 11021003, 11021008, 11101027, 11101028, 11122033, 11122034, 11152041, 11152042, 11152043, 11152044, 11152045, 11152046, 11152047, 11152048, 11152049, 11152050, 11152051, 11152052, 11152053, 11186054, 11186055, 11186056, 11186061, 11186062, 11186063, 11186064, 11186066, 11213071, 11213072, 11213073, 11213074, 11213075, 11213076, 11213078, 11213079, 11213080, 11213084, 11213085, 11236087, 11236089, 11236090, 11236091, 11236092, 11236093, 11250094, 11250095, 11250096, 11250098, 11250099, 11250100, 11250101, 11250103, 11250105, 11250106, 11250108, 11250109, 11276111, 11276112, 11276113, 11276114, 11276121, 11276122, 11276123, 11276124, 11276125, 11276127, 11276128, 11276129, 11306130, 11306131, 11306132, 11306133, 11306135, 11306136, 11306137, 11306141, 11306142, 11306144, 11306145, 11306146, 11306147, 11306148, 11306149, 11339151, 11339152, 11339153, 11339156, 11339157, 11339158, 11339159, 11339162, 11339163, 11339164, 12016001, 12016002, 12016003, 12016004, 12038011, 12038013, 12038014, 12038015, 12038016, 12038018, 12045021, 12045022, 12045023, 12045026, 12045027, 12045028, 12045029, 12045030, 12053034, 12053035, 12053036, 12053038, 12053039, 12053040, 12061041, 12061042, 12061043, 12061045, 12061048, 12061050, 12069052, 12069055, 12069056, 12069057, 12069058, 12069059, 12069060, 12079061, 12079062, 12079063, 12079064, 12079065, 12079066, 12079068, 12079069, 12079070, 12086071, 12086072, 12086073, 12086075, 12086076, 12086077, 12086078, 12086079, 12086080, 12088081, 12088082, 12088083, 12088084, 12088085, 12088086, 12088087, 12088088, 12088089, 12107091, 12107092, 12107093, 12107094, 12107095, 12107097, 12107098, 12111108, 12111109, 12111110, 12111111, 12111112, 12122056, 12128118, 12128119, 12128122, 12128123, 12128124, 12136125, 12136126, 12136127, 12136128, 12136134, 12136135, 12136136, 12142137, 12142138, 12142141, 12142142, 12145143, 12145144, 12145145, 12145146, 12145147, 12145148, 12145149, 12145150, 12145151, 12145152, 12145153, 12145154, 12156155, 12156156, 12156158, 12156160, 12156161, 12156162, 12156163, 12156164, 12156165,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Verigene Processor SP with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested. recalled?

Yes. FDA published recall Z-2776-2020 on August 19, 2020. Reported hazard: An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail