SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA
Published August 9, 2017 · Recall initiated: June 29, 2017
Summary
FDA records medical device Class II recall Z-2785-2017, published August 9, 2017 and initiated June 29, 2017. Recalling firm: Sorin Group USA, Inc.; Status: Terminated; Reported quantity: 524 units; Lot: Lot Numbers: 1620700048, 1621600063, 1623700039, 1624500038, 1625900047, 1626500084, 1627800066, 1628500062, 1630000040, 1630500055, 1631400035, 1633300052, 1635400049, 1700600038, 1702500069, 1704500039, 1706600107, 1707400082, 1708600072, 1710800102, 1711700034, 1713600045, 1715000038, 1717000046. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sorin Group USA, Inc.
- Distributor
- Nationwide
- Lot
- Lot Numbers: 1620700048, 1621600063, 1623700039, 1624500038, 1625900047, 1626500084, 1627800066, 1628500062, 1630000040, 1630500055, 1631400035, 1633300052, 1635400049, 1700600038, 1702500069, 1704500039, 1706600107, 1707400082, 1708600072, 1710800102, 1711700034, 1713600045, 1715000038, 1717000046
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASorin Group Aortic Arch Cannula, 7 mm x 10 in x 3/8 in, Rx Only, Sterile Product Usage: The Aortic ArchAugust 16, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDANON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, REF 627376401, 1 EA, NOT FOR CLINICAL USEAugust 9, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDABiosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (largeAugust 9, 2017
Questions about this recall
Was SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA recalled?
Yes. FDA published recall Z-2785-2017 on August 9, 2017. Reported hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone