Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US, SKU G4 PRO US G4 PRO Black US - Refurb V, SKU TG0002428-1V10 Pro Charger Single Unit, SKU PRO-CHARGER-SINGLE
Published August 12, 2026 · Recall initiated: June 23, 2026
Summary
FDA records medical device Class II recall Z-2835-2026, published August 12, 2026 and initiated June 23, 2026. Recalling firm: Therabody, Inc; Status: Ongoing; Reported quantity: 584224; Lot: UDI-DI 00810036050135 Charger GTIN 810036051378 All lots are recalled. Production ended in 2022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and overheating.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Therabody, Inc
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UDI-DI 00810036050135 Charger GTIN 810036051378 All lots are recalled. Production ended in 2022.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAREMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance DataAugust 12, 2026
- FDAFORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affectedAugust 12, 2026
- FDAHalyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH,August 12, 2026
- FDAKingfisher Compact Individual First Aid Kit. REF Number: 76656. Containing the affected One-piece GuedelAugust 12, 2026
- FDAHalyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMCAugust 12, 2026
- FDARadial Jaw 4 Standard Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 SCAugust 12, 2026
Questions about this recall
Was Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US, SKU G4 PRO US G4 PRO Black US - Refurb V, SKU TG0002428-1V10 Pro Charger Single Unit, SKU PRO-CHARGER-SINGLE recalled?
Yes. FDA published recall Z-2835-2026 on August 12, 2026. Reported hazard: Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and overheating.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail