FDAFDA · Food and drugs·Recall number Z-2835-2026

Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US, SKU G4 PRO US G4 PRO Black US - Refurb V, SKU TG0002428-1V10 Pro Charger Single Unit, SKU PRO-CHARGER-SINGLE

Published August 12, 2026 · Recall initiated: June 23, 2026

Summary

FDA records medical device Class II recall Z-2835-2026, published August 12, 2026 and initiated June 23, 2026. Recalling firm: Therabody, Inc; Status: Ongoing; Reported quantity: 584224; Lot: UDI-DI 00810036050135 Charger GTIN 810036051378 All lots are recalled. Production ended in 2022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and overheating.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Therabody, Inc
Distributor
Worldwide - US Nationwide distribution.
Lot
UDI-DI 00810036050135 Charger GTIN 810036051378 All lots are recalled. Production ended in 2022.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US, SKU G4 PRO US G4 PRO Black US - Refurb V, SKU TG0002428-1V10 Pro Charger Single Unit, SKU PRO-CHARGER-SINGLE recalled?

Yes. FDA published recall Z-2835-2026 on August 12, 2026. Reported hazard: Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and overheating.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail