DigitalDiagnost 4.1 High Performance (Stitching Patient Support) 712031
Published August 29, 2018 · Recall initiated: June 27, 2018
Summary
FDA records medical device Class II recall Z-2843-2018, published August 29, 2018 and initiated June 27, 2018. Recalling firm: Philips Medical Systems Gmbh, DMC; Status: Terminated; Lot: Serial Numbers: SN15000196 SN15000180 SN15000282 SN15000281 SN15000192 SN15000189. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems Gmbh, DMC
- Distributor
- US Nationwide; International to 66 countries
- Lot
- Serial Numbers: SN15000196 SN15000180 SN15000282 SN15000281 SN15000192 SN15000189
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDABuckyDiagnost Ceiling System (Stitching Patient Support) 704031August 29, 2018
- FDADigitalDiagnost Dual Detector (Stitching Patient Support) 712022August 29, 2018
- FDADigitalDiagnost Single Detector (Stitching Patient Support) 712020August 29, 2018
- FDAEasyDiagnost Eleva DRF (Stitching Patient Support) 706032August 29, 2018
- FDADigitalDiagnost 4 Flex / Value (Stitching Patient Support) 712028August 29, 2018
- FDAbioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMSAugust 29, 2018
Questions about this recall
Was DigitalDiagnost 4.1 High Performance (Stitching Patient Support) 712031 recalled?
Yes. FDA published recall Z-2843-2018 on August 29, 2018. Reported hazard: The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter