Back to search
FDAFDA · Food and drugs·Recall number Z-2868-2020

ALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 8®

Published September 2, 2020 · Recall initiated: July 22, 2020

Summary

FDA records medical device Class II recall Z-2868-2020, published September 2, 2020 and initiated July 22, 2020. Recalling firm: Globus Medical, Inc.; Status: Terminated; Reported quantity: N/A; Lot: Part #: 1124.1215; Lot # / UDI #: BAX112XE / (01)00889095069914(10)BAX112XE; BAX250LE / (01)00889095069914(10)BAX250LE. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Specific lots of ALTERA® implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Globus Medical, Inc.
Distributor
Worldwide Distribution
Lot
Part #: 1124.1215; Lot # / UDI #: BAX112XE / (01)00889095069914(10)BAX112XE; BAX250LE / (01)00889095069914(10)BAX250LE
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 8® recalled?

Yes. FDA published recall Z-2868-2020 on September 2, 2020. Reported hazard: Specific lots of ALTERA® implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

What should I do if I have this product?

Voluntary: Firm initiated · Letter