Back to search
FDAFDA · Food and drugs·Recall number Z-2878-2020

VITROS XT3400 Chemistry System, Product code 6844458

Published September 2, 2020 · Recall initiated: July 20, 2020

Summary

FDA records medical device Class II recall Z-2878-2020, published September 2, 2020 and initiated July 20, 2020. Recalling firm: Ortho Clinical Diagnostics; Status: Terminated; Reported quantity: pending; Lot: Serial numbers J34500122 J34500231. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho Clinical Diagnostics
Distributor
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Lot
Serial numbers J34500122 J34500231
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITROS XT3400 Chemistry System, Product code 6844458 recalled?

Yes. FDA published recall Z-2878-2020 on September 2, 2020. Reported hazard: When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter