INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Published August 12, 2026 · Recall initiated: June 17, 2026
Summary
FDA records medical device Class I recall Z-2879-2026, published August 12, 2026 and initiated June 17, 2026. Recalling firm: Boston Scientific Neuromodulation Corporation; Status: Ongoing; Reported quantity: 1079; Lot: Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Neuromodulation Corporation
- Distributor
- US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina
- Lot
- Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the NaluNovember 4, 2026
- FDAGE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring systemAugust 12, 2026
- FDASyva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319ULAugust 12, 2026
- FDAIntracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESSJuly 15, 2026
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180June 17, 2026
- FDABoston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF:June 17, 2026
Questions about this recall
Was INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems. recalled?
Yes. FDA published recall Z-2879-2026 on August 12, 2026. Reported hazard: Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
What should I do if I have this product?
Voluntary: Firm initiated · Letter