Back to search
FDAFDA · Food and drugs·Recall number Z-2879-2026

INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.

Published August 12, 2026 · Recall initiated: June 17, 2026

Summary

FDA records medical device Class I recall Z-2879-2026, published August 12, 2026 and initiated June 17, 2026. Recalling firm: Boston Scientific Neuromodulation Corporation; Status: Ongoing; Reported quantity: 1079; Lot: Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Neuromodulation Corporation
Distributor
US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina
Lot
Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems. recalled?

Yes. FDA published recall Z-2879-2026 on August 12, 2026. Reported hazard: Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

What should I do if I have this product?

Voluntary: Firm initiated · Letter