Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Published October 5, 2016 · Recall initiated: December 29, 2009
Summary
FDA records medical device Class II recall Z-2888-2016, published October 5, 2016 and initiated December 29, 2009. Recalling firm: Spiegelberg Gmbh & Co. KG; Status: Terminated; Reported quantity: 22 units; Lot: Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393; Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Spiegelberg Gmbh & Co. KG
- Distributor
- Nationwide Distribution to PA only
- Lot
- Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393; Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAVarithena (polidocanol injectable foam) Administration Pack For Use With Varithena, Rx only, Distributed byOctober 12, 2016
- FDAMethylcobalamin 5000 mcg/mL Injection, packaged in 1, 2, 3, 4, and 5 mL vials, IM, Rx, The Apothecary ShoppeOctober 12, 2016
- FDATRUEread Blood Glucose Test Strips (Glucose Oxidase) are used with TRUEresult and TRUE2go meters. TRUEreadOctober 5, 2016
- FDABruxZir Shaded, Model No. 70-1167-NOW0001, 70-1167-NOW0002, 70-1167-NOW0003, 70-1167-NOW0005, 70-1167-NOW0006October 5, 2016
- FDAJakes Candy Gummies Worms. o Product label(s): ***Jakes Candy *** GUMMIES WORMS *** Net Wt. 1.1 oz. *** Exp.October 5, 2016
- FDATriana Family Brioche Donuts Bavarian Cream Net Wt. 0.75 LBS (340 g) 12 -24 Jabez St., Newark, NJ 07106 UPCOctober 5, 2016
Questions about this recall
Was Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head. recalled?
Yes. FDA published recall Z-2888-2016 on October 5, 2016. Reported hazard: In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
What should I do if I have this product?
Voluntary: Firm initiated · Letter