BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.
Published September 16, 2020 · Recall initiated: August 4, 2020
Summary
FDA records medical device Class I recall Z-2939-2020, published September 16, 2020 and initiated August 4, 2020. Recalling firm: CareFusion 303, Inc.; Status: Ongoing; Reported quantity: 248,519; Lot: All serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- Worldwide: US* (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY); United Arab Emirates (AE); Australia (AU); Canada (CA); Hungary (HU); Israel (IL); Japan (JP)* Kuwait (KW); New Zealand (NZ); Philippines (PH); Qatar (QA); Saudi Arabia (SA); Singapore (SG); Taiwan (TW); Turkey (TR)*. *There are 159 U.S. government and/or military consignees including two located in Japan (JP) and Turkey (TR).
- Lot
- All serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinetJune 10, 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinetJune 10, 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinetJune 10, 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinetJune 10, 2026
- FDASINGLES Carbide US-No. FG556; UDI +E0HM26000313140091; STERILE R; Size 009, Length 3.8; 2025-03-18 - ProductSeptember 16, 2020
- FDATalladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No.September 16, 2020
Questions about this recall
Was BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump. recalled?
Yes. FDA published recall Z-2939-2020 on September 16, 2020. Reported hazard: Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.
What should I do if I have this product?
Voluntary: Firm initiated · Letter