Back to search
FDAFDA · Food and drugs·Recall number Z-2960-2026

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Published August 26, 2026 · Recall initiated: July 8, 2026

Summary

FDA records medical device Class II recall Z-2960-2026, published August 26, 2026 and initiated July 8, 2026. Recalling firm: Tobii Dynavox Llc; Status: Ongoing; Lot: UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Tobii Dynavox Llc
Distributor
US Nationwide distribution.
Lot
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799 recalled?

Yes. FDA published recall Z-2960-2026 on August 26, 2026. Reported hazard: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

What should I do if I have this product?

Voluntary: Firm initiated · Letter