Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288

Published September 12, 2018 · Recall initiated: June 28, 2018

Summary

FDA records medical device Class II recall Z-2961-2018, published September 12, 2018 and initiated June 28, 2018. Recalling firm: Bayer Medical Care, Inc.; Status: Terminated; Reported quantity: 1; Lot: Serial Number: 4061630901. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Bayer Medical Care, Inc.
Distributor
The devices were distributed to the following US states: OH, TX, and WV.
Lot
Serial Number: 4061630901
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288 recalled?

Yes. FDA published recall Z-2961-2018 on September 12, 2018. Reported hazard: The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

What should I do if I have this product?

Voluntary: Firm initiated · Letter