Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 020661 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay
Published September 23, 2020 · Recall initiated: August 7, 2020
Summary
FDA records medical device Class II recall Z-2968-2020, published September 23, 2020 and initiated August 7, 2020. Recalling firm: Tosoh Bioscience Inc; Status: Terminated; Reported quantity: 39 boxes; Lot: Lot Numbers: J803507, JX03508, JZ03509, A203510, A403511. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tosoh Bioscience Inc
- Distributor
- Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.
- Lot
- Lot Numbers: J803507, JX03508, JZ03509, A203510, A403511
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone inJuly 22, 2026
- FDAZenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment ofSeptember 23, 2020
- FDAImactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with theSeptember 23, 2020
- FDAST AIA-PACK ACTH (ACTH: Adrenocorticotropic Hormone) is designed for IN VITRO DIAGNOSTIC USE ONLY for theJanuary 2, 2019
- FDAAIA-PACK Folate (FOL) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of folateJanuary 2, 2019
- FDAST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurementJanuary 2, 2019
Questions about this recall
Was Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 020661 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay recalled?
Yes. FDA published recall Z-2968-2020 on September 23, 2020. Reported hazard: Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
What should I do if I have this product?
Voluntary: Firm initiated · Letter