Back to search
FDAFDA · Food and drugs·Recall number Z-2974-2026

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Published August 26, 2026 · Recall initiated: July 15, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide distribution - US Nationwide.
Lot
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card