Back to search
FDAFDA · Food and drugs·Recall number Z-2977-2024

N-LHP-928, HeNe Laser System, 35 mW

Published September 18, 2024 · Recall initiated: March 1, 2024

Summary

FDA records medical device Class II recall Z-2977-2024, published September 18, 2024 and initiated March 1, 2024. Recalling firm: PACIFIC LASERTECH, LLC; Status: Ongoing; Reported quantity: 32; Lot: N-LHP-928. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

What to do

FDA Mandated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
PACIFIC LASERTECH, LLC
Distributor
US Nationwide Distribution
Lot
N-LHP-928
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was N-LHP-928, HeNe Laser System, 35 mW recalled?

Yes. FDA published recall Z-2977-2024 on September 18, 2024. Reported hazard: These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

What should I do if I have this product?

FDA Mandated · Letter