Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.
Published September 2, 2026 · Recall initiated: July 13, 2026
Summary
FDA records medical device Class II recall Z-2991-2026, published September 2, 2026 and initiated July 13, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 20 units; Lot: All lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- Worldwide distribution - US Nationwide and the country of Panama.
- Lot
- All lot codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAPhilips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.September 2, 2026
- FDAMedline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline ProductSeptember 2, 2026
Questions about this recall
Was Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier. recalled?
Yes. FDA published recall Z-2991-2026 on September 2, 2026. Reported hazard: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit