Boston Scientific Wallstent Endoprosthesis Tracheobronchial Transhepatic Biliary Self-Expanding Stent, 12MM X 90MM | 75CM L, resterilized.
Published September 19, 2018 · Recall initiated: March 7, 2018
Summary
FDA records medical device Class II recall Z-3035-2018, published September 19, 2018 and initiated March 7, 2018. Recalling firm: SPS Sterilization, Inc; Status: Terminated; Reported quantity: 2 units total; Lot: Code Number 40213-R; Lot Number 18030192; Serial Number 21408032017-10, 21408032017-11. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SPS Sterilization, Inc
- Distributor
- US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
- Lot
- Code Number 40213-R; Lot Number 18030192; Serial Number 21408032017-10, 21408032017-11
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEdwards Lifesciences Avid Dual Stage Venous Drainage Cannula, 29/37F (9.6/12.3MM) X 37CM (14.5"),September 19, 2018
- FDAEdwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 34F (11.3MM) X 40CM (16"), resterilized.September 19, 2018
- FDAEdwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 38FRX40CM, resterilized.September 19, 2018
- FDAVITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only.September 19, 2018
- FDAAffixus® Hip Fracture Nail 125® 9 mm x 180 mm, Item Number: 814309180September 19, 2018
- FDAAffixus® Hip Fracture Nail Right 125® 9 mm x 320 mm, Item Number: 814309320September 19, 2018
Questions about this recall
Was Boston Scientific Wallstent Endoprosthesis Tracheobronchial Transhepatic Biliary Self-Expanding Stent, 12MM X 90MM | 75CM L, resterilized. recalled?
Yes. FDA published recall Z-3035-2018 on September 19, 2018. Reported hazard: The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
What should I do if I have this product?
Voluntary: Firm initiated · Letter