STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test
Published September 9, 2026 · Recall initiated: July 29, 2026
Summary
FDA records medical device Class II recall Z-3047-2026, published September 9, 2026 and initiated July 29, 2026. Recalling firm: Diagnostica Stago, Inc.; Status: Ongoing; Reported quantity: 5220 units; Lot: UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Diagnostica Stago, Inc.
- Distributor
- US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.
- Lot
- UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHalyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2)September 9, 2026
- FDAMedline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A-September 9, 2026
- FDASTA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use withSeptember 3, 2025
- FDADiagnostica Stago STA N®oplastine® CI, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombinSeptember 19, 2018
- FDANeoplastine® Cl Plus ¿ (ref. 00375) Product Usage: Manual or automated determination of the prothrombin timeMarch 21, 2018
- FDASTA - Neoplastine® Cl Plus ¿ (ref. 00606) Product Usage: Manual or automated determination of the prothrombinMarch 21, 2018
Questions about this recall
Was STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test recalled?
Yes. FDA published recall Z-3047-2026 on September 9, 2026. Reported hazard: A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail