Medline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375C
Published September 9, 2026 · Recall initiated: June 12, 2026
Summary
FDA records medical device Class II recall Z-3053-2026, published September 9, 2026 and initiated June 12, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 300 kits; Lot: 1) DYNJ63375C, UDI-DI: 10193489851823(each), 40193489851824(case), Lot Number: 23LMC035. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution.
- Lot
- 1) DYNJ63375C, UDI-DI: 10193489851823(each), 40193489851824(case), Lot Number: 23LMC035
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAMedline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437DSeptember 9, 2026
- FDAMedline Convenience Kits: 1) IV START KIT, Catalog Number: DYNDV2520; 2) PERIPHERAL LAB BLOOD CULTURE KIT,September 9, 2026
Questions about this recall
Was Medline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375C recalled?
Yes. FDA published recall Z-3053-2026 on September 9, 2026. Reported hazard: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
What should I do if I have this product?
Voluntary: Firm initiated · Letter