Dimension Vista® Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Published September 18, 2024 · Recall initiated: July 30, 2024
Summary
FDA records medical device Class II recall Z-3065-2024, published September 18, 2024 and initiated July 30, 2024. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 5,371 units (4,394 US, 977 OUS); Lot: Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.
- Lot
- Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151September 18, 2024
- FDABrand Name: Vit-l-escence Product Name: Vit-l-escence Singles A3 Model/Catalog Number: 1203- Software Version:September 18, 2024
Questions about this recall
Was Dimension Vista® Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146; recalled?
Yes. FDA published recall Z-3065-2024 on September 18, 2024. Reported hazard: During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
What should I do if I have this product?
Voluntary: Firm initiated · Letter