FDAFDA · Food and drugs·Recall number Z-3067-2017
Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Published August 30, 2017 · Recall initiated: May 23, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Pentax of America Inc
- Distributor
- Nationwide
- Lot
- Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411, H110497, G110360, G110363, G110364, G110371, G110366, G110368, G110372, H110434, H110447, G110380, H110455, G110298, H110445, H110449, H110461, G110292, G110346, G110365, H110494, H110495, H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467, H110471, H110412, G110289, G110303, G110294, H110420, G110299, G110300, H110448, H110452, G110386, G110331, H110451, H110493, G110313, G110314, H110464, H110496, H110433, G110373, G110375, G110317, H110431, H110417, G110291, G110302, G110369, G110378, G110335, G110387, G110390, G110392, G110395, G110311, G110327, G110332, G110382, G110355, G110357, G110287, G110320, H110435, H110481, H110485, H110486, G110354, G110316, G110318, G110362, G110393, H110498, H110456, H110457, H110459, G110337, G110356, H110421, H110429, G110290, G110352, H110470, H110478, G110383, G110384, G110361.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States