Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

Published August 30, 2017 · Recall initiated: May 23, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Pentax of America Inc
Distributor
Nationwide
Lot
Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411, H110497, G110360, G110363, G110364, G110371, G110366, G110368, G110372, H110434, H110447, G110380, H110455, G110298, H110445, H110449, H110461, G110292, G110346, G110365, H110494, H110495, H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467, H110471, H110412, G110289, G110303, G110294, H110420, G110299, G110300, H110448, H110452, G110386, G110331, H110451, H110493, G110313, G110314, H110464, H110496, H110433, G110373, G110375, G110317, H110431, H110417, G110291, G110302, G110369, G110378, G110335, G110387, G110390, G110392, G110395, G110311, G110327, G110332, G110382, G110355, G110357, G110287, G110320, H110435, H110481, H110485, H110486, G110354, G110316, G110318, G110362, G110393, H110498, H110456, H110457, H110459, G110337, G110356, H110421, H110429, G110290, G110352, H110470, H110478, G110383, G110384, G110361.
Risk classification
Class II
Area (FDA)
device
Country
United States