FDAFDA · Food and drugs·Recall number Z-3071-2017

CADD(R) Medication Cassette Reservoir with clamp and female Luer. Non vented stopper included, REF 21-7002-24, Rx only, STERILE EO, Manufacturer Smiths Medical ASD, Inc.

Published September 6, 2017 · Recall initiated: June 5, 2017

Summary

FDA records medical device Class II recall Z-3071-2017, published September 6, 2017 and initiated June 5, 2017. Recalling firm: Smiths Medical ASD Inc.; Status: Terminated; Reported quantity: 17280 units; Lot: lot number 16X659. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The medication cassette reservoir, part number 21-7002-24, with lot number 16X659, may have been manufactured with the incorrect pressure plate and the tubing used on the cassette may have been routed incorrectly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Japan, China, US
Lot
lot number 16X659
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CADD(R) Medication Cassette Reservoir with clamp and female Luer. Non vented stopper included, REF 21-7002-24, Rx only, STERILE EO, Manufacturer Smiths Medical ASD, Inc. recalled?

Yes. FDA published recall Z-3071-2017 on September 6, 2017. Reported hazard: The medication cassette reservoir, part number 21-7002-24, with lot number 16X659, may have been manufactured with the incorrect pressure plate and the tubing used on the cassette may have been routed incorrectly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter