FDAFDA · Food and drugs·Recall number Z-3075-2017

VITEK® 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.

Published September 6, 2017 · Recall initiated: July 12, 2017

Summary

FDA records medical device Class II recall Z-3075-2017, published September 6, 2017 and initiated July 12, 2017. Recalling firm: Biomerieux Inc; Status: Terminated; Reported quantity: 1866 units; Lot: VITEK® 2 Compact 15 and Compact 30 instruments previously manufactured/installed with VITEK® 2 System software versions prior to 5.01, and currently being upgraded to version 8.01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Customers have reported that some VITEK® 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK® 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomerieux Inc
Distributor
Worldwide
Lot
VITEK® 2 Compact 15 and Compact 30 instruments previously manufactured/installed with VITEK® 2 System software versions prior to 5.01, and currently being upgraded to version 8.01.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITEK® 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast. recalled?

Yes. FDA published recall Z-3075-2017 on September 6, 2017. Reported hazard: Customers have reported that some VITEK® 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK® 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail