Back to search
FDAFDA · Food and drugs·Recall number Z-3075-2026

Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators PATIENT ADMIN KIT DYKA1384B DYKA1384C WELCOME KIT DYKS1206D

Published September 9, 2026 · Recall initiated: July 20, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that were recalled by the manufacturer because they may have an open or incomplete seal on the applicator packaging. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medline Industries, LP
Distributor
US Nationwide distribution.
Lot
DYKA1384B UDI-DI 10198459415753 Lot 25GBS617 DYKA1384C UDI-DI 10198459485206 Lot 25IBF898 DYKS1206D UDI-DI 10198459414794 Lots 25HBI181 25JBJ452 25LBJ223 26ABT372 26CBF527 26EBH733
Risk classification
Class II
Area (FDA)
device
Country
United States