FDAFDA · Food and drugs·Recall number Z-3105-2017
Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems
Published September 13, 2017 · Recall initiated: August 14, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Merge Healthcare, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Software versions 10.0, 10.0.1, 10.0.2, 10.0.3, 10.0.4, 10.0.5, 11.0, 11.0.1, 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1, 11.1.1.1, 11.1.1 Patch 2, 11.1.1 Patch 3, 11.1.2, 11.1.2 Patch 1, 11.1.2 Patch 2, 11.1.2 Patch 3, and 11.1.2 Patch 4. Only customers with one of the above listed software versions AND a license for Mammo Tracking are impacted by this recall.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States