FDAFDA · Food and drugs·Recall number Z-3107-2024
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack Number DYKE1558B; c) ENDO KIT, Pack Number DYKE1804A; d) VASCULAR-LF, Pack Number DYNJ900714C; e) MAJOR CARDIOVASC ADULT, Pack Number DYNJ905472B; f) AV FISTULA, Pack Number DYNJ908096A
Published September 18, 2024 · Recall initiated: June 27, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, Panama.
- Lot
- a) DYK1002174I: UDI/DI case 40195327332403, UDI/DI each 10195327332402; Lot Numbers: 24CLA959; b) DYKE1558B: UDI/DI case 40195327258512, UDI/DI each 10195327258511; Lot Numbers: 24DMA783, 24BMG878, 23LMG790; c) DYKE1804A: UDI/DI case 40195327023035, UDI/DI each 10195327023034; Lot Numbers: 24BMI437, 24AME311, 23LMC958; d) DYNJ900714C: UDI/DI case 40193489951364, UDI/DI each 10193489951363; Lot Numbers: 24FBC609; e) DYNJ905472B: UDI/DI case 40195327200573, UDI/DI each 10195327200572; Lot Numbers: 24BBS291; f) DYNJ908096A: UDI/DI case 40195327507061, UDI/DI each 10195327507060; Lot Numbers: 24ABX948, 24ABB080
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.