Back to search
FDAFDA · Food and drugs·Recall number Z-3109-2026

Halyard IV Start Kit Non-Sterile REF: KRIV26-02

Published September 16, 2026 · Recall initiated: July 14, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which lacks sterility assurance due to to wrinkles in the paper lidding which may breach the seal area.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
AVID Medical, Inc.
Distributor
US Nationwide distribution in the state of NE.
Lot
Lot Number:1658438 7/31/2026 UDI: 10809160460481
Risk classification
Class II
Area (FDA)
device
Country
United States