Back to search
FDAFDA · Food and drugs·Recall number Z-3113-2024

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMINOPLASTY W/LIPO SUCTION, Pack Number DYNJ909946; d) PLASTIC, Pack Number DYNJ909951

Published September 18, 2024 · Recall initiated: June 27, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Lot
a) DYNJ908104B: UDI/DI case 40195327507115, UDI/DI each 10195327507114, Lot Numbers: 23LBT869, 24BBQ730; b) DYNJ908106B: UDI/DI case 40195327507139, UDI/DI each 10195327507138, Lot Numbers: 24ABB087, 24CBM739, 24EBH234; c) DYNJ909946: UDI/DI case 40195327512652, UDI/DI each 10195327512651, Lot Numbers: 23LBD934, 24DBB806; d) DYNJ909951: UDI/DI case 40195327512904, UDI/DI each 10195327512903, Lot Numbers: 23LBD932, 24ABP740
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card