FDAFDA · Food and drugs·Recall number Z-3122-2024

FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W

Published September 18, 2024 · Recall initiated: August 7, 2024

Summary

FDA records medical device Class II recall Z-3122-2024, published September 18, 2024 and initiated August 7, 2024. Recalling firm: BioFire Diagnostics, LLC; Status: Ongoing; Reported quantity: 1001; Lot: FLM2-ASY-0001, UDI: 00815381020031, Serial Numbers: 2FA00610, 2FA00951, 2FA00964, 2FA01005, 2FA01115, 2FA01116, 2FA01128, 2FA01418, 2FA01653, 2FA01733, 2FA01771, 2FA01775, 2FA01878, 2FA02805, 2FA02977, 2FA02981, 2FA03194, 2FA03196, 2FA03237, 2FA03273, 2FA03303, 2FA03335, 2FA03343, 2FA06079, 2FA06473, 2FA07884, 2FA07987, 2FA07996, 2FA08386, 2FA01224, 2FAD848B, 2FA00229, 2FA00386, 2FA00396, 2FA00551, 2FA00558, 2FA01693, 2FA02188, 2FA02296, 2FA02347, 2FA02422, 2FA05086, 2FA05089, 2FA05099, 2FA05172, 2FA05186, 2FA05256, 2FA05273, 2FA05287, 2FA05300, 2FA05601, 2FA06055, 2FA06140, 2FA06236, 2FA06297, 2FA06655, 2FA06867, 2FA07093, 2FA07173, 2FA07175, 2FA07194, 2FA07260, 2FA07312, 2FA07728, 2FA07848, 2FA07849, 2FA07856, 2FA07859, 2FA07860, 2FA07867, 2FA07870, 2FA07892, 2FA07916, 2FA07933, 2FA07976, 2FA07978, 2FA08054, 2FA08145, 2FA08336, 2FA08559, 2FA08579, 2FA08700, 2FA08728, 2FA08785, 2FA09867, FA2461, FA2464, FA2893, FA3170, FA3566, 2FA09893, 2FA00673, 2FA00983, 2FA01178, 2FA01250, 2FA02300, 2FA02823, 2FA03096, 2FA03115, 2FA03119, 2FA03133, 2FA03490, 2FA05263, 2FA05602, 2FA05950, 2FA06200, 2FA06274, 2FA06484, 2FA06645, 2FA07047, 2FA07253, 2FA07334, 2FA07388, 2FA07445, 2FA07446, 2FA07452, 2FA07471, 2FA07488, 2FA07682, 2FA07705, 2FA06431, 2FA09166, 2FAA364A, 2FAB4CAC, 2FAE02D5, 2FA08587, 2FA07899, 2FA01903, 2FA02860, 2FA05962, 2FA06659, 2FA06740, 2FA06746, 2FA06811, 2FA06905, 2FA06912, 2FA06954, 2FA06955, 2FA07084, 2FA07222, 2FA07225, 2FA07267, 2FA07350, 2FA07423, 2FA07450, 2FA07603, 2FA07812, 2FA07822, 2FA07964, 2FA08017, 2FA08045, 2FA08188, 2FA2AD2E, 2FA3331D, 2FAC0903, 2FA03026, 2FA05845, 2FA06670, 2FA06724, 2FA07795, 2FA08558, 2FA08782, 2FA09350, 2FA09371, 2FA09376. FLM2-ASY-0001-W, UDI: 00815381029058, Serial Numbers: 2FA00046, 2FA00049, 2FA00050, 2FA00053, 2FA00060, 2FA00066, 2FA00076, 2FA00079, 2FA00090, 2FA00093, 2FA00116, 2FA00117, 2FA00131, 2FA00139, 2FA00142, 2FA00150, 2FA00157, 2FA00159, 2FA00167, 2FA00169, 2FA00172, 2FA00174, 2FA00176, 2FA00185, 2FA00186, 2FA00187, 2FA00209, 2FA00210, 2FA00217, 2FA00220, 2FA00221, 2FA00275, 2FA00278, 2FA00295, 2FA00298, 2FA00299, 2FA00308, 2FA00333, 2FA00340, 2FA00343, 2FA00344, 2FA00346, 2FA00348, 2FA00359, 2FA00360, 2FA00365, 2FA00366, 2FA00369, 2FA00372, 2FA00378, 2FA00390, 2FA00403, 2FA00405, 2FA00413, 2FA00415, 2FA00424, 2FA00435, 2FA00436, 2FA00440, 2FA00442, 2FA00452, 2FA00453, 2FA00455, 2FA00456, 2FA00459, 2FA00483, 2FA00498, 2FA00500, 2FA00521, 2FA00529, 2FA00540, 2FA00542, 2FA00543, 2FA00544, 2FA00561, 2FA00570, 2FA00582, 2FA00611, 2FA00618, 2FA00620, 2FA00633, 2FA00658, 2FA00676, 2FA00678, 2FA00682, 2FA00684, 2FA00692, 2FA00695, 2FA00715, 2FA00716, 2FA00731, 2FA00738, 2FA00760, 2FA00795, 2FA00813, 2FA00838, 2FA00842, 2FA00847, 2FA00864, 2FA00881, 2FA00882, 2FA00886, 2FA00887, 2FA00893, 2FA00904, 2FA00908, 2FA00909, 2FA00922, 2FA00942, 2FA00948, 2FA00952, 2FA00960, 2FA00965, 2FA00970, 2FA00971, 2FA00980, 2FA00984, 2FA01001, 2FA01028, 2FA01035, 2FA01041, 2FA01043, 2FA01047, 2FA01051, 2FA01056, 2FA01068, 2FA01088, 2FA01106, 2FA01121, 2FA01122, 2FA01133, 2FA01151, 2FA01155, 2FA01168, 2FA01175, 2FA01216, 2FA01226, 2FA01230, 2FA01232, 2FA01233, 2FA01242, 2FA01259, 2FA01270, 2FA01274, 2FA01275, 2FA01288, 2FA01290, 2FA01292, 2FA01294, 2FA01302, 2FA01312, 2FA01325, 2FA01335, 2FA01336, 2FA01344, 2FA01351, 2FA01358, 2FA01366, 2FA01370, 2FA01378, 2FA01380, 2FA01389, 2FA01390, 2FA01393, 2FA01408, 2FA01419, 2FA01429, 2FA01435, 2FA01444, 2FA01449, 2FA01462, 2FA01471, 2FA01473, 2FA01476, 2FA01491, 2FA01520, 2FA01543, 2FA01555, 2FA01559, 2FA01568, 2FA01571, 2FA01574, 2FA01579, 2FA01605, 2FA01612, 2FA01650, 2FA01651, 2FA01670, 2FA01688, 2FA01731, 2FA01740, 2FA01748, 2FA01787, 2FA01804, 2FA01810, 2FA01814, 2FA01820, 2FA01824, 2FA01847, 2FA01858, 2FA01860, 2FA01876, 2FA01888, 2FA01929, 2FA01932, 2FA02005, 2FA02006, 2FA02009, 2FA02015, 2FA02018, 2FA02023, 2FA02029, 2FA02033, 2FA02035, 2FA02040, 2FA02086, 2FA02098, 2FA02110, 2FA02125, 2FA02134, 2FA02140, 2FA02147, 2FA02199, 2FA02201, 2FA02233, 2FA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
BioFire Diagnostics, LLC
Distributor
Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea, Canada, Singapore distributors, Canada, Canada, China, China, Singapore, Canada, China, Mexico, Japan, Chile, Chile, Singapore, Mexico, Hong King, China, Chile, Singapore, Philippines, Korea, Chile, Korea, Australia, Greece.
Lot
FLM2-ASY-0001, UDI: 00815381020031, Serial Numbers: 2FA00610, 2FA00951, 2FA00964, 2FA01005, 2FA01115, 2FA01116, 2FA01128, 2FA01418, 2FA01653, 2FA01733, 2FA01771, 2FA01775, 2FA01878, 2FA02805, 2FA02977, 2FA02981, 2FA03194, 2FA03196, 2FA03237, 2FA03273, 2FA03303, 2FA03335, 2FA03343, 2FA06079, 2FA06473, 2FA07884, 2FA07987, 2FA07996, 2FA08386, 2FA01224, 2FAD848B, 2FA00229, 2FA00386, 2FA00396, 2FA00551, 2FA00558, 2FA01693, 2FA02188, 2FA02296, 2FA02347, 2FA02422, 2FA05086, 2FA05089, 2FA05099, 2FA05172, 2FA05186, 2FA05256, 2FA05273, 2FA05287, 2FA05300, 2FA05601, 2FA06055, 2FA06140, 2FA06236, 2FA06297, 2FA06655, 2FA06867, 2FA07093, 2FA07173, 2FA07175, 2FA07194, 2FA07260, 2FA07312, 2FA07728, 2FA07848, 2FA07849, 2FA07856, 2FA07859, 2FA07860, 2FA07867, 2FA07870, 2FA07892, 2FA07916, 2FA07933, 2FA07976, 2FA07978, 2FA08054, 2FA08145, 2FA08336, 2FA08559, 2FA08579, 2FA08700, 2FA08728, 2FA08785, 2FA09867, FA2461, FA2464, FA2893, FA3170, FA3566, 2FA09893, 2FA00673, 2FA00983, 2FA01178, 2FA01250, 2FA02300, 2FA02823, 2FA03096, 2FA03115, 2FA03119, 2FA03133, 2FA03490, 2FA05263, 2FA05602, 2FA05950, 2FA06200, 2FA06274, 2FA06484, 2FA06645, 2FA07047, 2FA07253, 2FA07334, 2FA07388, 2FA07445, 2FA07446, 2FA07452, 2FA07471, 2FA07488, 2FA07682, 2FA07705, 2FA06431, 2FA09166, 2FAA364A, 2FAB4CAC, 2FAE02D5, 2FA08587, 2FA07899, 2FA01903, 2FA02860, 2FA05962, 2FA06659, 2FA06740, 2FA06746, 2FA06811, 2FA06905, 2FA06912, 2FA06954, 2FA06955, 2FA07084, 2FA07222, 2FA07225, 2FA07267, 2FA07350, 2FA07423, 2FA07450, 2FA07603, 2FA07812, 2FA07822, 2FA07964, 2FA08017, 2FA08045, 2FA08188, 2FA2AD2E, 2FA3331D, 2FAC0903, 2FA03026, 2FA05845, 2FA06670, 2FA06724, 2FA07795, 2FA08558, 2FA08782, 2FA09350, 2FA09371, 2FA09376. FLM2-ASY-0001-W, UDI: 00815381029058, Serial Numbers: 2FA00046, 2FA00049, 2FA00050, 2FA00053, 2FA00060, 2FA00066, 2FA00076, 2FA00079, 2FA00090, 2FA00093, 2FA00116, 2FA00117, 2FA00131, 2FA00139, 2FA00142, 2FA00150, 2FA00157, 2FA00159, 2FA00167, 2FA00169, 2FA00172, 2FA00174, 2FA00176, 2FA00185, 2FA00186, 2FA00187, 2FA00209, 2FA00210, 2FA00217, 2FA00220, 2FA00221, 2FA00275, 2FA00278, 2FA00295, 2FA00298, 2FA00299, 2FA00308, 2FA00333, 2FA00340, 2FA00343, 2FA00344, 2FA00346, 2FA00348, 2FA00359, 2FA00360, 2FA00365, 2FA00366, 2FA00369, 2FA00372, 2FA00378, 2FA00390, 2FA00403, 2FA00405, 2FA00413, 2FA00415, 2FA00424, 2FA00435, 2FA00436, 2FA00440, 2FA00442, 2FA00452, 2FA00453, 2FA00455, 2FA00456, 2FA00459, 2FA00483, 2FA00498, 2FA00500, 2FA00521, 2FA00529, 2FA00540, 2FA00542, 2FA00543, 2FA00544, 2FA00561, 2FA00570, 2FA00582, 2FA00611, 2FA00618, 2FA00620, 2FA00633, 2FA00658, 2FA00676, 2FA00678, 2FA00682, 2FA00684, 2FA00692, 2FA00695, 2FA00715, 2FA00716, 2FA00731, 2FA00738, 2FA00760, 2FA00795, 2FA00813, 2FA00838, 2FA00842, 2FA00847, 2FA00864, 2FA00881, 2FA00882, 2FA00886, 2FA00887, 2FA00893, 2FA00904, 2FA00908, 2FA00909, 2FA00922, 2FA00942, 2FA00948, 2FA00952, 2FA00960, 2FA00965, 2FA00970, 2FA00971, 2FA00980, 2FA00984, 2FA01001, 2FA01028, 2FA01035, 2FA01041, 2FA01043, 2FA01047, 2FA01051, 2FA01056, 2FA01068, 2FA01088, 2FA01106, 2FA01121, 2FA01122, 2FA01133, 2FA01151, 2FA01155, 2FA01168, 2FA01175, 2FA01216, 2FA01226, 2FA01230, 2FA01232, 2FA01233, 2FA01242, 2FA01259, 2FA01270, 2FA01274, 2FA01275, 2FA01288, 2FA01290, 2FA01292, 2FA01294, 2FA01302, 2FA01312, 2FA01325, 2FA01335, 2FA01336, 2FA01344, 2FA01351, 2FA01358, 2FA01366, 2FA01370, 2FA01378, 2FA01380, 2FA01389, 2FA01390, 2FA01393, 2FA01408, 2FA01419, 2FA01429, 2FA01435, 2FA01444, 2FA01449, 2FA01462, 2FA01471, 2FA01473, 2FA01476, 2FA01491, 2FA01520, 2FA01543, 2FA01555, 2FA01559, 2FA01568, 2FA01571, 2FA01574, 2FA01579, 2FA01605, 2FA01612, 2FA01650, 2FA01651, 2FA01670, 2FA01688, 2FA01731, 2FA01740, 2FA01748, 2FA01787, 2FA01804, 2FA01810, 2FA01814, 2FA01820, 2FA01824, 2FA01847, 2FA01858, 2FA01860, 2FA01876, 2FA01888, 2FA01929, 2FA01932, 2FA02005, 2FA02006, 2FA02009, 2FA02015, 2FA02018, 2FA02023, 2FA02029, 2FA02033, 2FA02035, 2FA02040, 2FA02086, 2FA02098, 2FA02110, 2FA02125, 2FA02134, 2FA02140, 2FA02147, 2FA02199, 2FA02201, 2FA02233, 2FA
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W recalled?

Yes. FDA published recall Z-3122-2024 on September 18, 2024. Reported hazard: Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

What should I do if I have this product?

Voluntary: Firm initiated · Letter