StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.
Published September 23, 2026 · Recall initiated: July 29, 2026
Summary
FDA records medical device Class I recall Z-3130-2026, published September 23, 2026 and initiated July 29, 2026. Recalling firm: Nova Biomedical Corporation; Status: Ongoing; Reported quantity: 1,858 units; Lot: Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nova Biomedical Corporation
- Distributor
- Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.
- Lot
- Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAStatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.September 23, 2026
- FDAStatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.September 23, 2026
- FDAStatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.September 23, 2026
- FDAStatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.September 23, 2026
- FDAMedline Convenience Kits: 1) DYNDB3086D, Catalog Number: DYNDB3086D; 2) DYNDB3100, Catalog Number: DYNDB3100;September 23, 2026
- FDAMedline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008NSeptember 23, 2026
Questions about this recall
Was StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system. recalled?
Yes. FDA published recall Z-3130-2026 on September 23, 2026. Reported hazard: A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail