FDAFDA · Food and drugs·Recall number Z-3132-2017
FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable
Published September 20, 2017 · Recall initiated: May 15, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.
What to do
Voluntary: Firm initiated · Visit
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- US Distribution to one customer in Missouri.
- Lot
- Lot/Batch No. (Exp Date): SPM01975 (09/05/2044), SPM00890 (09/27/2043), SPM01616 (06/28/2044)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States