FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable

Published September 20, 2017 · Recall initiated: May 15, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.

What to do

Voluntary: Firm initiated · Visit

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
US Distribution to one customer in Missouri.
Lot
Lot/Batch No. (Exp Date): SPM01975 (09/05/2044), SPM00890 (09/27/2043), SPM01616 (06/28/2044)
Risk classification
Class II
Area (FDA)
device
Country
United States