INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.
Published September 30, 2026 · Recall initiated: August 11, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- CooperSurgical, Inc.
- Distributor
- Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
- Lot
- 1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624. 2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057. 3. Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623. 4. Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per bo…September 30, 2026
- FDAHalyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Mo…September 30, 2026
- FDALemon Alfredo Fettuccine packed in a yellow plastic sealed bag with pink and black lettering; label reads in …September 30, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/…September 30, 2026
- FDAFromagerie Coop de Mean Cabricharme and Belgium Cabricharme Shelf life: 30 days, all dates through and includ…September 30, 2026
- FDABrand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Versi…December 25, 2024
- FDABrand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Descriptio…December 25, 2024
- FDABrand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23055 Software Version: N/A …December 25, 2024
Questions about this recall
Was INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715. recalled?
Yes. FDA published recall Z-3140-2026 on September 30, 2026. Reported hazard: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.