Back to search
FDAFDA · Food and drugs·Recall number Z-3143-2026

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Published September 23, 2026 · Recall initiated: July 30, 2026

Summary

FDA records medical device Class II recall Z-3143-2026, published September 23, 2026 and initiated July 30, 2026. Recalling firm: GE Medical Systems Information Technologies Inc; Status: Ongoing; Reported quantity: 29,570 (23,632 US; 5,938 OUS); Lot: 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Medical Systems Information Technologies Inc
Distributor
Worldwide - US Nationwide distribution.
Lot
5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV recalled?

Yes. FDA published recall Z-3143-2026 on September 23, 2026. Reported hazard: Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

What should I do if I have this product?

Voluntary: Firm initiated · Letter