Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
Published September 23, 2026 · Recall initiated: July 30, 2026
Summary
FDA records medical device Class II recall Z-3143-2026, published September 23, 2026 and initiated July 30, 2026. Recalling firm: GE Medical Systems Information Technologies Inc; Status: Ongoing; Reported quantity: 29,570 (23,632 US; 5,938 OUS); Lot: 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems Information Technologies Inc
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips PIC iX Essentials REF:867093September 23, 2026
- FDAPercutaneous Insertion Kit (PIK), Guidewire 100 cm, Model Numbers: 701053068 (USA), 701047384 (OUS);September 23, 2026
- FDAPORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209September 2, 2026
- FDAFRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 5700225September 2, 2026
- FDAPORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592September 2, 2026
- FDAPORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353September 2, 2026
Questions about this recall
Was Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV recalled?
Yes. FDA published recall Z-3143-2026 on September 23, 2026. Reported hazard: Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
What should I do if I have this product?
Voluntary: Firm initiated · Letter