Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US.

Published September 23, 2026 · Recall initiated: August 12, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Abiomed, Inc.
Distributor
Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
Lot
1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA; UDI-DI: 00813502010022; Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715, IC1769, IC1790, IC1791, IC1911, IC1912, IC1913, IC2032, IC2033, IC2034, IC2035, IC2369, IC2370, IC2686, IC2687, IC3661, IC3662, IC3802, IC3803, IC4944, IC4945, IC5034, IC5050, IC5680, IC5821, IC5824, IC6082, IC6083, IC6633, IC6634, IC6635, IC6636, IC7694, IC7695, IC7716, IC8283, IC9413, IC9414, IC9415, IC9416; 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU; UDI-DI: 00813502011289; Serial Numbers: IC1347, IC1349, IC6845, IC1285, IC1416, IC2765, IC2766, IC3021, IC3022, IC3718, IC3719, IC3720, IC3776, IC5558, IC5559, IC1578, IC1585, IC1902, IC1903, IC1905, IC2678, IC2679, IC2855, IC2856, IC3053, IC3054, IC3329, IC3671, IC3672, IC3830, IC3831, IC4033, IC4837, IC5086, IC5169, IC5171, IC5560, IC5681, IC5682, IC5683, IC6064, IC6081, IC6149, IC6270, IC6272, IC6437, IC6439, IC6441, IC6443, IC6810, IC6812, IC6813, IC6814, IC7261, IC7262, IC7393, IC7394, IC7576, IC7577, IC7646, IC1825, IC1826, IC1870, IC1876, IC4153, IC4175, IC4657, IC4658, IC5476, IC5477, IC5518, IC5520, IC7263, IC7264, IC1011, IC1490, IC1584, IC1590, IC1592, IC2175, IC2176, IC2857, IC2996, IC2997, IC3291, IC3296, IC3317, IC3318, IC3486, IC3488, IC3698, IC3702, IC3724, IC3778, IC3864, IC3870, IC3890, IC3891, IC4151, IC4152, IC4323, IC4324, IC4471, IC4501, IC4681, IC4682, IC4851, IC4892, IC4933, IC4934, IC4973, IC4984, IC4986, IC4987, IC4988, IC5007, IC5048, IC5175, IC5176, IC5195, IC5421, IC5422, IC5433, IC5434, IC5441, IC5442, IC5460, IC5461, IC5463, IC5464, IC5494, IC5495, IC5496, IC5519, IC5611, IC5684, IC5687, IC5688, IC5714, IC5722, IC5818, IC6065, IC6534, IC6535, IC6536, IC6582, IC6587, IC6588, IC6589, IC6590, IC6688, IC6689, IC6690, IC6691, IC6692, IC6693, IC6694, IC6695, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6741, IC6742, IC6876, IC6877, IC6878, IC6879, IC7388, IC7389, IC7390, IC7392, IC7647, IC7648, IC7649, IC7650, IC7651, IC7652, IC1284, IC2676, IC2677, IC2710, IC2827, IC2828, IC2989, IC3018, IC3020, IC3471, IC5775, IC5777, IC5779, IC5780, IC2099, IC2205, IC2220, IC2654, IC2655, IC2823, IC2824, IC2825, IC2826, IC3668, IC3670, IC3868, IC3869, IC4099, IC4100, IC4209, IC4218, IC4325, IC4326, IC4569, IC4611, IC4624, IC4627, IC4931, IC4932, IC5490, IC5491, IC5492, IC5493, IC6205, IC6275, IC6331, IC6432, IC6434, IC6435, IC6880, IC7407, IC7421, IC7455, IC7675, IC7686, IC7691, IC7692, IC7693, IC7717, IC7720, IC7721, IC7748, IC7809, IC3131, IC3132, IC1549, IC1580, IC2234, IC2236, IC2469, IC2652, IC2653, IC3069, IC3070, IC3311, IC3331, IC3332, IC3657, IC3669, IC3777, IC3828, IC3829, IC4150, IC1396, IC1397, IC1413, IC1415, IC1546, IC1548, IC1724, IC1725, IC1767, IC1768, IC2235, IC2239, IC2483, IC2486, IC2836, IC2838, IC3050, IC3673, IC3696, IC3697, IC4565, IC4839, IC5755, IC5756, IC6487, IC6488, IC7667, IC7674, IC7689, IC7690, IC3301, IC3302, IC1301, IC1302, IC1304, IC1305, IC1217, IC1276, IC1384, IC1386, IC1395, IC1522, IC1523, IC1547, IC1577, IC1579, IC1594, IC1676, IC1684, IC1699, IC1700, IC1722, IC1723, IC1764, IC1821, IC1824, IC1886, IC1974, IC2101, IC2102, IC2329, IC2862, IC2873, IC2874, IC2875, IC2876, IC2877, IC4002, IC4012, IC4013, IC4014, IC4015, IC4464, IC4465, IC4470, IC4566, IC4597, IC4598, IC4609, IC4623, IC4634, IC4635, IC4636, IC4645, IC4656, IC4833, IC4834, IC4835, IC4836, IC4850, IC4852, IC4928, IC4929, IC4930, IC5051, IC5053, IC5057, IC5075, IC5685, IC6579, IC6591, IC6592, IC6593, IC6594, IC6595, IC6708, IC6709, IC6711, IC6715, IC6716, IC6717, IC6804, IC6805, IC6806, IC6807, IC6808, IC6809, IC7098, IC7099, IC7100, IC7258, IC7259, IC7260, IC7580, IC7613, IC7614, IC7615, IC7687, IC7688, IC3817, IC3818, IC5008, IC5009, IC5046, IC5047, IC6223, IC6224, IC7286, IC7378, IC7087, IC7092, IC1265, IC1275, IC1280, IC1283, IC1288, IC1344, IC1345, IC1350, IC1383, IC1411, IC1653, IC1726, IC1761, IC1775, IC1780, IC1881, IC1901, IC1904, IC1
Risk classification
Class I
Area (FDA)
device
Country
United States