FDAFDA · Food and drugs·Recall number Z-3148-2026

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US.

Published September 23, 2026 · Recall initiated: August 12, 2026

Summary

FDA records medical device Class I recall Z-3148-2026, published September 23, 2026 and initiated August 12, 2026. Recalling firm: Abiomed, Inc.; Status: Ongoing; Reported quantity: 5232 units (3733 US, 1499 OUS) 36 3; Lot: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA; UDI-DI: 00813502010022; Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715, IC1769, IC1790, IC1791, IC1911, IC1912, IC1913, IC2032, IC2033, IC2034, IC2035, IC2369, IC2370, IC2686, IC2687, IC3661, IC3662, IC3802, IC3803, IC4944, IC4945, IC5034, IC5050, IC5680, IC5821, IC5824, IC6082, IC6083, IC6633, IC6634, IC6635, IC6636, IC7694, IC7695, IC7716, IC8283, IC9413, IC9414, IC9415, IC9416; 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU; UDI-DI: 00813502011289; Serial Numbers: IC1347, IC1349, IC6845, IC1285, IC1416, IC2765, IC2766, IC3021, IC3022, IC3718, IC3719, IC3720, IC3776, IC5558, IC5559, IC1578, IC1585, IC1902, IC1903, IC1905, IC2678, IC2679, IC2855, IC2856, IC3053, IC3054, IC3329, IC3671, IC3672, IC3830, IC3831, IC4033, IC4837, IC5086, IC5169, IC5171, IC5560, IC5681, IC5682, IC5683, IC6064, IC6081, IC6149, IC6270, IC6272, IC6437, IC6439, IC6441, IC6443, IC6810, IC6812, IC6813, IC6814, IC7261, IC7262, IC7393, IC7394, IC7576, IC7577, IC7646, IC1825, IC1826, IC1870, IC1876, IC4153, IC4175, IC4657, IC4658, IC5476, IC5477, IC5518, IC5520, IC7263, IC7264, IC1011, IC1490, IC1584, IC1590, IC1592, IC2175, IC2176, IC2857, IC2996, IC2997, IC3291, IC3296, IC3317, IC3318, IC3486, IC3488, IC3698, IC3702, IC3724, IC3778, IC3864, IC3870, IC3890, IC3891, IC4151, IC4152, IC4323, IC4324, IC4471, IC4501, IC4681, IC4682, IC4851, IC4892, IC4933, IC4934, IC4973, IC4984, IC4986, IC4987, IC4988, IC5007, IC5048, IC5175, IC5176, IC5195, IC5421, IC5422, IC5433, IC5434, IC5441, IC5442, IC5460, IC5461, IC5463, IC5464, IC5494, IC5495, IC5496, IC5519, IC5611, IC5684, IC5687, IC5688, IC5714, IC5722, IC5818, IC6065, IC6534, IC6535, IC6536, IC6582, IC6587, IC6588, IC6589, IC6590, IC6688, IC6689, IC6690, IC6691, IC6692, IC6693, IC6694, IC6695, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6741, IC6742, IC6876, IC6877, IC6878, IC6879, IC7388, IC7389, IC7390, IC7392, IC7647, IC7648, IC7649, IC7650, IC7651, IC7652, IC1284, IC2676, IC2677, IC2710, IC2827, IC2828, IC2989, IC3018, IC3020, IC3471, IC5775, IC5777, IC5779, IC5780, IC2099, IC2205, IC2220, IC2654, IC2655, IC2823, IC2824, IC2825, IC2826, IC3668, IC3670, IC3868, IC3869, IC4099, IC4100, IC4209, IC4218, IC4325, IC4326, IC4569, IC4611, IC4624, IC4627, IC4931, IC4932, IC5490, IC5491, IC5492, IC5493, IC6205, IC6275, IC6331, IC6432, IC6434, IC6435, IC6880, IC7407, IC7421, IC7455, IC7675, IC7686, IC7691, IC7692, IC7693, IC7717, IC7720, IC7721, IC7748, IC7809, IC3131, IC3132, IC1549, IC1580, IC2234, IC2236, IC2469, IC2652, IC2653, IC3069, IC3070, IC3311, IC3331, IC3332, IC3657, IC3669, IC3777, IC3828, IC3829, IC4150, IC1396, IC1397, IC1413, IC1415, IC1546, IC1548, IC1724, IC1725, IC1767, IC1768, IC2235, IC2239, IC2483, IC2486, IC2836, IC2838, IC3050, IC3673, IC3696, IC3697, IC4565, IC4839, IC5755, IC5756, IC6487, IC6488, IC7667, IC7674, IC7689, IC7690, IC3301, IC3302, IC1301, IC1302, IC1304, IC1305, IC1217, IC1276, IC1384, IC1386, IC1395, IC1522, IC1523, IC1547, IC1577, IC1579, IC1594, IC1676, IC1684, IC1699, IC1700, IC1722, IC1723, IC1764, IC1821, IC1824, IC1886, IC1974, IC2101, IC2102, IC2329, IC2862, IC2873, IC2874, IC2875, IC2876, IC2877, IC4002, IC4012, IC4013, IC4014, IC4015, IC4464, IC4465, IC4470, IC4566, IC4597, IC4598, IC4609, IC4623, IC4634, IC4635, IC4636, IC4645, IC4656, IC4833, IC4834, IC4835, IC4836, IC4850, IC4852, IC4928, IC4929, IC4930, IC5051, IC5053, IC5057, IC5075, IC5685, IC6579, IC6591, IC6592, IC6593, IC6594, IC6595, IC6708, IC6709, IC6711, IC6715, IC6716, IC6717, IC6804, IC6805, IC6806, IC6807, IC6808, IC6809, IC7098, IC7099, IC7100, IC7258, IC7259, IC7260, IC7580, IC7613, IC7614, IC7615, IC7687, IC7688, IC3817, IC3818, IC5008, IC5009, IC5046, IC5047, IC6223, IC6224, IC7286, IC7378, IC7087, IC7092, IC1265, IC1275, IC1280, IC1283, IC1288, IC1344, IC1345, IC1350, IC1383, IC1411, IC1653, IC1726, IC1761, IC1775, IC1780, IC1881, IC1901, IC1904, IC1. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abiomed, Inc.
Distributor
Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
Lot
1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA; UDI-DI: 00813502010022; Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715, IC1769, IC1790, IC1791, IC1911, IC1912, IC1913, IC2032, IC2033, IC2034, IC2035, IC2369, IC2370, IC2686, IC2687, IC3661, IC3662, IC3802, IC3803, IC4944, IC4945, IC5034, IC5050, IC5680, IC5821, IC5824, IC6082, IC6083, IC6633, IC6634, IC6635, IC6636, IC7694, IC7695, IC7716, IC8283, IC9413, IC9414, IC9415, IC9416; 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU; UDI-DI: 00813502011289; Serial Numbers: IC1347, IC1349, IC6845, IC1285, IC1416, IC2765, IC2766, IC3021, IC3022, IC3718, IC3719, IC3720, IC3776, IC5558, IC5559, IC1578, IC1585, IC1902, IC1903, IC1905, IC2678, IC2679, IC2855, IC2856, IC3053, IC3054, IC3329, IC3671, IC3672, IC3830, IC3831, IC4033, IC4837, IC5086, IC5169, IC5171, IC5560, IC5681, IC5682, IC5683, IC6064, IC6081, IC6149, IC6270, IC6272, IC6437, IC6439, IC6441, IC6443, IC6810, IC6812, IC6813, IC6814, IC7261, IC7262, IC7393, IC7394, IC7576, IC7577, IC7646, IC1825, IC1826, IC1870, IC1876, IC4153, IC4175, IC4657, IC4658, IC5476, IC5477, IC5518, IC5520, IC7263, IC7264, IC1011, IC1490, IC1584, IC1590, IC1592, IC2175, IC2176, IC2857, IC2996, IC2997, IC3291, IC3296, IC3317, IC3318, IC3486, IC3488, IC3698, IC3702, IC3724, IC3778, IC3864, IC3870, IC3890, IC3891, IC4151, IC4152, IC4323, IC4324, IC4471, IC4501, IC4681, IC4682, IC4851, IC4892, IC4933, IC4934, IC4973, IC4984, IC4986, IC4987, IC4988, IC5007, IC5048, IC5175, IC5176, IC5195, IC5421, IC5422, IC5433, IC5434, IC5441, IC5442, IC5460, IC5461, IC5463, IC5464, IC5494, IC5495, IC5496, IC5519, IC5611, IC5684, IC5687, IC5688, IC5714, IC5722, IC5818, IC6065, IC6534, IC6535, IC6536, IC6582, IC6587, IC6588, IC6589, IC6590, IC6688, IC6689, IC6690, IC6691, IC6692, IC6693, IC6694, IC6695, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6741, IC6742, IC6876, IC6877, IC6878, IC6879, IC7388, IC7389, IC7390, IC7392, IC7647, IC7648, IC7649, IC7650, IC7651, IC7652, IC1284, IC2676, IC2677, IC2710, IC2827, IC2828, IC2989, IC3018, IC3020, IC3471, IC5775, IC5777, IC5779, IC5780, IC2099, IC2205, IC2220, IC2654, IC2655, IC2823, IC2824, IC2825, IC2826, IC3668, IC3670, IC3868, IC3869, IC4099, IC4100, IC4209, IC4218, IC4325, IC4326, IC4569, IC4611, IC4624, IC4627, IC4931, IC4932, IC5490, IC5491, IC5492, IC5493, IC6205, IC6275, IC6331, IC6432, IC6434, IC6435, IC6880, IC7407, IC7421, IC7455, IC7675, IC7686, IC7691, IC7692, IC7693, IC7717, IC7720, IC7721, IC7748, IC7809, IC3131, IC3132, IC1549, IC1580, IC2234, IC2236, IC2469, IC2652, IC2653, IC3069, IC3070, IC3311, IC3331, IC3332, IC3657, IC3669, IC3777, IC3828, IC3829, IC4150, IC1396, IC1397, IC1413, IC1415, IC1546, IC1548, IC1724, IC1725, IC1767, IC1768, IC2235, IC2239, IC2483, IC2486, IC2836, IC2838, IC3050, IC3673, IC3696, IC3697, IC4565, IC4839, IC5755, IC5756, IC6487, IC6488, IC7667, IC7674, IC7689, IC7690, IC3301, IC3302, IC1301, IC1302, IC1304, IC1305, IC1217, IC1276, IC1384, IC1386, IC1395, IC1522, IC1523, IC1547, IC1577, IC1579, IC1594, IC1676, IC1684, IC1699, IC1700, IC1722, IC1723, IC1764, IC1821, IC1824, IC1886, IC1974, IC2101, IC2102, IC2329, IC2862, IC2873, IC2874, IC2875, IC2876, IC2877, IC4002, IC4012, IC4013, IC4014, IC4015, IC4464, IC4465, IC4470, IC4566, IC4597, IC4598, IC4609, IC4623, IC4634, IC4635, IC4636, IC4645, IC4656, IC4833, IC4834, IC4835, IC4836, IC4850, IC4852, IC4928, IC4929, IC4930, IC5051, IC5053, IC5057, IC5075, IC5685, IC6579, IC6591, IC6592, IC6593, IC6594, IC6595, IC6708, IC6709, IC6711, IC6715, IC6716, IC6717, IC6804, IC6805, IC6806, IC6807, IC6808, IC6809, IC7098, IC7099, IC7100, IC7258, IC7259, IC7260, IC7580, IC7613, IC7614, IC7615, IC7687, IC7688, IC3817, IC3818, IC5008, IC5009, IC5046, IC5047, IC6223, IC6224, IC7286, IC7378, IC7087, IC7092, IC1265, IC1275, IC1280, IC1283, IC1288, IC1344, IC1345, IC1350, IC1383, IC1411, IC1653, IC1726, IC1761, IC1775, IC1780, IC1881, IC1901, IC1904, IC1
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US. recalled?

Yes. FDA published recall Z-3148-2026 on September 23, 2026. Reported hazard: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter