Back to search
FDAFDA · Food and drugs·Recall number Z-3152-2024

2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT

Published September 25, 2024 · Recall initiated: July 11, 2024

Summary

FDA records medical device Class II recall Z-3152-2024, published September 25, 2024 and initiated July 11, 2024. Recalling firm: Medtronic Inc.; Status: Ongoing; Reported quantity: 18788 units; Lot: 2090 Programmers with serial number prefixes PKK0 and PKK1: a) GTIN: 00613994136480, 00613994136497, 00613994136503, 00613994136510, 00613994136527, 00613994136534, 00613994136565, 00613994136596, 00613994136619, 00613994136626, 00613994175076, 00613994175083, 00613994175090, 00613994175106, 00613994175120, 00613994175137, 00613994175144, 00613994175151, 00613994175168, 00613994175175, 00613994175182, 00613994175205, 00613994175212, 00613994175229, 00613994175236, 00613994175243, 00613994175250, 00613994175267, 00613994175281, 00613994175298, 00613994175304, 00613994175311, 00613994175328, 00613994175335, 00613994175342, 00613994175366, 00613994175373, 00613994175380, 00613994175397, 00613994175403, 00613994175410, 00613994175434, 00613994175465, 00613994175489, 00613994175540, 00613994175588, 00613994175595, 00613994203182, 00613994360922, 00613994434883, 00613994434890, 00613994434944, 00613994435002, 00613994435255, 00613994435316, 00613994580863, 00613994580894, 00613994580931, 00613994580962, 00613994581013, 00613994581051, 00613994581075, 00613994581105, 00613994581150, 00613994581167, 00613994581174, 00613994581181, 00613994581204, 00613994581211, 00613994581228, 00613994581235, 00613994581242, 00613994581549, 00613994581570, 00613994581587, 00613994581594, 00613994581624, 00643169440128, 00643169440166, 00643169440173, 00643169440197, 00643169440203, 00643169440937, 00643169537750, 00643169537774, 00643169537811, 00643169537828, 00643169537835, 00643169537842, 00643169537859, 00643169537866, 00643169537880, 00643169537910, 00643169537927, 00643169538016, 00643169629004, 00643169629011, 00643169629028, 00643169629035, 00643169629042, 00643169629066, 00721902246227, 00721902246234, 00721902246241, 00721902246265, 00721902246289, 00721902246302, 00721902246326, 00721902246333; b) GTIN: 00721902171222; c) GTIN: 00721902171079; d) GTIN: 00721902824692; e) GTIN: 00721902059100; f) GTIN: 00721902171154; g) GTIN: 00721902246357; h) GTIN: 00721902246364; i) GTIN: 00721902480201; j) GTIN: 00721902246159, 00721902246173; k) GTIN: 00721902246203; l) GTIN: 00721902841903; m) GTIN: 00721902246272; n) GTIN: 00613994127631, 00613994435354, 00613994581549, 00885074301440, 00885074310596, 00885074310633, 00885074310657, 00885074310930, 00885074312866, 00885074312873, 00885074313047, 00885074313108, 00885074313115, 00885074313139, 00885074313146, 00885074313153, 00885074313160; o) GTIN: 00885074313191; p) GTIN: 00885074313177; q) GTIN: 00885074313184; r) GTIN: 00885074314921; s) GTIN: 00885074313009; t) GTIN: 00885074357416; u) GTIN: 00885074313122. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Inc.
Distributor
Worldwide distribution.
Lot
2090 Programmers with serial number prefixes PKK0 and PKK1: a) GTIN: 00613994136480, 00613994136497, 00613994136503, 00613994136510, 00613994136527, 00613994136534, 00613994136565, 00613994136596, 00613994136619, 00613994136626, 00613994175076, 00613994175083, 00613994175090, 00613994175106, 00613994175120, 00613994175137, 00613994175144, 00613994175151, 00613994175168, 00613994175175, 00613994175182, 00613994175205, 00613994175212, 00613994175229, 00613994175236, 00613994175243, 00613994175250, 00613994175267, 00613994175281, 00613994175298, 00613994175304, 00613994175311, 00613994175328, 00613994175335, 00613994175342, 00613994175366, 00613994175373, 00613994175380, 00613994175397, 00613994175403, 00613994175410, 00613994175434, 00613994175465, 00613994175489, 00613994175540, 00613994175588, 00613994175595, 00613994203182, 00613994360922, 00613994434883, 00613994434890, 00613994434944, 00613994435002, 00613994435255, 00613994435316, 00613994580863, 00613994580894, 00613994580931, 00613994580962, 00613994581013, 00613994581051, 00613994581075, 00613994581105, 00613994581150, 00613994581167, 00613994581174, 00613994581181, 00613994581204, 00613994581211, 00613994581228, 00613994581235, 00613994581242, 00613994581549, 00613994581570, 00613994581587, 00613994581594, 00613994581624, 00643169440128, 00643169440166, 00643169440173, 00643169440197, 00643169440203, 00643169440937, 00643169537750, 00643169537774, 00643169537811, 00643169537828, 00643169537835, 00643169537842, 00643169537859, 00643169537866, 00643169537880, 00643169537910, 00643169537927, 00643169538016, 00643169629004, 00643169629011, 00643169629028, 00643169629035, 00643169629042, 00643169629066, 00721902246227, 00721902246234, 00721902246241, 00721902246265, 00721902246289, 00721902246302, 00721902246326, 00721902246333; b) GTIN: 00721902171222; c) GTIN: 00721902171079; d) GTIN: 00721902824692; e) GTIN: 00721902059100; f) GTIN: 00721902171154; g) GTIN: 00721902246357; h) GTIN: 00721902246364; i) GTIN: 00721902480201; j) GTIN: 00721902246159, 00721902246173; k) GTIN: 00721902246203; l) GTIN: 00721902841903; m) GTIN: 00721902246272; n) GTIN: 00613994127631, 00613994435354, 00613994581549, 00885074301440, 00885074310596, 00885074310633, 00885074310657, 00885074310930, 00885074312866, 00885074312873, 00885074313047, 00885074313108, 00885074313115, 00885074313139, 00885074313146, 00885074313153, 00885074313160; o) GTIN: 00885074313191; p) GTIN: 00885074313177; q) GTIN: 00885074313184; r) GTIN: 00885074314921; s) GTIN: 00885074313009; t) GTIN: 00885074357416; u) GTIN: 00885074313122
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT recalled?

Yes. FDA published recall Z-3152-2024 on September 25, 2024. Reported hazard: Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit