System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12
Published September 25, 2024 · Recall initiated: August 22, 2024
Summary
FDA records medical device Class II recall Z-3165-2024, published September 25, 2024 and initiated August 22, 2024. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 7091 units; Lot: Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C®te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R®union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen
- Lot
- Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAPhilips Azurion 5 M12. Model Number: 722227.November 4, 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.November 4, 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.November 4, 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.November 4, 2026
- FDA1. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2001, 3" x 6", sterile; 2. Tech MedicalSeptember 25, 2024
- FDASystem Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 ORSeptember 25, 2024
Questions about this recall
Was System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12 recalled?
Yes. FDA published recall Z-3165-2024 on September 25, 2024. Reported hazard: Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter