Back to search
FDAFDA · Food and drugs·Recall number Z-3169-2017

ECT Internal Fracture Fixation Drill Bits 2.7 mm Diameter (.106 Inch), Item Number/EDI 00231802706, Nonsterile. orthopedic surgical instrument.

Published October 4, 2017 · Recall initiated: May 22, 2017

Summary

FDA records medical device Class II recall Z-3169-2017, published October 4, 2017 and initiated May 22, 2017. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 1,286 units; Lot: Lot Numbers: 63556699, 63556701, 63593922, 63616886, 63630279, 63641603, 63358167, 63372864, 63372865, 63377514, 63377516, 63415331, 63426482, 63431757, 63457850, 63457853, 63457856, 63469884, 63474434, 63479002, 63479004, 63479006 & 63524437. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer Biomet, Inc.
Distributor
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
Lot
Lot Numbers: 63556699, 63556701, 63593922, 63616886, 63630279, 63641603, 63358167, 63372864, 63372865, 63377514, 63377516, 63415331, 63426482, 63431757, 63457850, 63457853, 63457856, 63469884, 63474434, 63479002, 63479004, 63479006 & 63524437.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ECT Internal Fracture Fixation Drill Bits 2.7 mm Diameter (.106 Inch), Item Number/EDI 00231802706, Nonsterile. orthopedic surgical instrument. recalled?

Yes. FDA published recall Z-3169-2017 on October 4, 2017. Reported hazard: Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit