STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095W
Published September 25, 2024 · Recall initiated: August 15, 2024
Summary
FDA records medical device Class II recall Z-3177-2024, published September 25, 2024 and initiated August 15, 2024. Recalling firm: B Braun Medical Inc; Status: Ongoing; Reported quantity: 142336 units; Lot: UDI-DI 04046955348800; All Unexpired Lot Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B Braun Medical Inc
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI 04046955348800; All Unexpired Lot Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAB Braun SAFSITE Injection Site, RV1000NC SAFSITE VALVE W/O EXTRA CAP. Catalog Number: 415068.October 28, 2026
- FDABrand Name: Extension Set Product Name: DISCOFIX FOUR WAY STOPCOCK Model/Catalog Number: 456025September 9, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: CESK RIVERSIDE HEALTH SYSTEM Model/Catalog Number: 552126June 10, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: NES1725KFXL SPINAL EPID CLOSED TIP CATH Model/Catalog Number:June 10, 2026
- FDASTREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and aSeptember 25, 2024
- FDASTREAMLINE BLOODLINE SET FOR DIALOG-Tubing to be used to transfer blood between a patient and a haemodialyserSeptember 25, 2024
Questions about this recall
Was STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095W recalled?
Yes. FDA published recall Z-3177-2024 on September 25, 2024. Reported hazard: Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
What should I do if I have this product?
Voluntary: Firm initiated · Letter